Depth Adjustment
7.20.1
Depth Adjustment
Observe the angle of the footplate for reinstallation.
The settings for positioning the footplates on the half-clamps
may vary for each footplate.
1.
Remove the two flat screws
A
, washers
B
and locknuts
C
that
secure footplate
D
to the half clamp
E
.
2.
Move footplate to one of four mounting positions.
If desired depth is still not obtained, rotate the half
clamp 180° on the footplate hinge
F
.
3.
Retighten the two flat screws, washers and locknuts.
7.20.2
Angle Adjustment
Observe the angle of the footplate for reinstallation.
The settings for positioning the footplates on the half-clamps
may vary for each footplate.
1.
Loosen, but DO NOT remove, the two flat screws
A
, washer
B
and locknuts
C
that secure the footplate
D
to the half clamp
E
.
2.
Position the footplate to the necessary angle to accommodate
the user.
3.
Retighten the two flat screws, washers and locknuts.
7.20.3
Perpendicular and/or Inversion/Eversion
Adjustment
It is not necessary to remove the footplate to perform this
adjustment.
1.
Insert a flathead screwdriver through the half clamp
E
on the
footplate.
2.
Slowly turn nylon adjustment screw
G
in or out until footplate is
perpendicular to the footrest assembly or the desired inversion
or eversion is obtained.
1143192-L-03