background image

Dolomite Step Up®

3 Product Overview

3.1 Intended Use

The walking aid is intended to support a person’s ability to
stand and walk through gait training.

The walking aid is for indoor use only. It may be used in
wet areas, but not under pouring water (e.g. shower). Step
Up Electric must only be used in wet areas, when it’s not
connected to the mains.

The weight of the user must not exceed the maximum user
weight as stated in chapter

8.1 Dimensions and Weights,

page 17

and on the product label.

Contraindications

There are no contraindications known.

3.2 Product Description

Step Up is a height adjustable walking table available in
different versions.

Step Up Manual

Walking table, height adjustment operated manually, with:

Foot operated brake on all four wheels (model 1521677),

Foot operated brake on the front wheels and hand
operated / foot operated brake and directional lock on
the rear wheels (model 1585401).

Step up Hydraulic

Walking table, height adjustment operated by a gas spring,
with:

Foot operated brake on all four wheels (model 1521678),

Foot operated brake on the front wheels and hand
operated brake on the rear wheels (model 1536121),

Foot operated brake on the front wheels and hand
operated / foot operated brake and directional lock on
the rear wheels (model 1587608).

Step Up Electric

Walking table, height adjustment operated by an attenuator,
with:

Foot operated brake on all four wheels (models 1521679
and 1579540),

Foot operated brake on the front wheels and hand
operated brake on the rear wheels (model 1536122),

Foot operated brake on the front wheels and hand
operated / foot operated brake and directional lock on
the rear wheels (model 1587609).

Models 1521679 and 1536122 have an external charger.
Models 1579540 and 1587609 have an internal charger.

3.3 Scope of Delivery

Walking table, fully assembled

Control box and hand control (all Step Up Electric
models)

External charger (only Step Up Electric models 1521679
and 1536122)

User manual

3.4 Labels and Symbols on Product

Identification Label

X  XXXXXX    XXXXXX

Invaca re  REA AB

Växjövägen  303

SE-34 3 71 Diö

SWED EN

xxx kg

XXXX XXX XX XX

XXX XXX

XX XXXXX XXXX

 XX XX

XXXX XXXX XX X

 XXXX

XXXX

xx-xx cm

YYMXX000000

YYMXX000000

YYYY-MM

DOLOMITE Step Up xxx

XXXXXXXX

H

M

N

R

X

X

X

X

X

B

C

E

A

F

G

H

D

The label contains all important technical information
regarding your product.

A

Product name

F

Maximum user weight

B

Reference number

G

Maximum width

C

Date of manufacture

H

Electrical data*

D

Manufacturer address

E

Serial number

* Only on label of Step Up Electric models

Other Symbols on Label

European
Conformity

Read user manual

Medical Device

Maximum width

Reference
number

Maximum user
weight

Serial number

Type B Applied
Part*

Date of
manufacture

CLASS II
equipment*

Manufacturer

WEEE conform*

Indoor use

* Only on label of Step Up Electric models

Other Symbols on Product

Refer to the user manual.

For further electrical information for the Step Up
Electric, check label on the motor.

6

1532567-G

Содержание 1521677

Страница 1: ...40 1585401 1587608 1587609 en Walking table User Manual This manual MUST be given to the user of the product BEFORE using this product this manual MUST be read and saved for future reference 1521677 1...

Страница 2: ...whole or in part is prohibited without prior written permission from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unles...

Страница 3: ...3 Height adjustment Step Up Electric 10 4 9 Rising to a standing position 10 5 Accessories 11 5 1 Accessories General 11 5 2 Hand Brakes 11 5 2 1 Adjusting the brakes 11 5 3 Hand Brakes for Models wit...

Страница 4: ...inor or slight injury if it is not avoided IMPORTANT Indicates a hazardous situation that could result in damage to property if it is not avoided Tips and Recommendations Gives useful tips recommendat...

Страница 5: ...ating the walking table in the presence of children and pets Take care when moving over obstacles e g doorways when using the walking table with the accessory standing plate The wheels must be locked...

Страница 6: ...on all four wheels models 1521679 and 1579540 Foot operated brake on the front wheels and hand operated brake on the rear wheels model 1536122 Foot operated brake on the front wheels and hand operated...

Страница 7: ...onal Lock A directional lock is used to lock the wheel in a straight position in order to ease movement straight ahead A 1 Press down brake pedal A one step in order to activate directional lock 2 To...

Страница 8: ...angle of 25 between shoulder and wrist position gives the best conditions to enable the torso to remain upright The position is pre set and is maintained regardless of width setting 25 Set the armres...

Страница 9: ...s through the most narrow passage in the premises A B 1 Lay the walking table down 2 Remove screws A 3 Adjust the frame B to the required width on one side 4 Reinsert the screws and tighten with a min...

Страница 10: ...y distributed weight affects the product s stability 125 mm 215 mm Step Up Electric and Step Up Hydraulic can support the user when rising to a standing position The help of an attendant is recommende...

Страница 11: ...ke handle handles A slowly towards handles The more you press the higher the braking force is Engaging Parking Brake The parking brake sets the brake into a locked parking position 1 A To engage parki...

Страница 12: ...owly towards handles The more you press the higher the braking force is Engaging Parking Brake The parking brake sets the brake into a locked parking position 1 A B Press brake handle handles A toward...

Страница 13: ...bottom of the oxygen cylinder must be at least 8 cm from the ground 5 8 IV pole A The IV pole can be mounted on the pillar A and makes it possible to attach a drip bag on the product The Oxygen tank h...

Страница 14: ...tand Assist A B The Stand Assist accessory is available for users who need support while standing and consists of a standing plate A knee support B and a harness With the Stand Assist accessory the wa...

Страница 15: ...e in the cleaning instructions Never use a solvent cellulose thinner acetone etc that changes the structure of the plastic or dissolves the attached labels Always make sure that the product is complet...

Страница 16: ...rding to legislation DO NOT dispose of batteries in normal household waste Batteries MUST be taken to a proper disposal site The return is required by law and free of charge Do only dispose discharged...

Страница 17: ...mm with standard wheel 190 mm with optional 100 mm wheel Maximum user weight 150 kg Maximum load in basket and tray 5 kg Maximum load on the oxygen tank holder 11 kg Maximum load on the IV pole 4 kg P...

Страница 18: ...d to the mains Make sure the walking table is dry before connection it to the mains Operation Storage Transportation Temperature 5 C to 40 C 10 C to 50 C 10 C to 50 C Relative humidity 20 to 100 20 to...

Страница 19: ...ied below The user of the product should assure that it is used in such an environment Immunity test IEC 60601 1 2 test level Compliance level Electromagnetic environment guidance Electrostatic discha...

Страница 20: ...strength in the location in which this product is used exceeds the applicable RF compliance level above the product should be observed to verify normal operation If abnormal performance is observed a...

Страница 21: ...MHz the separation distance for the higher frequency range applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures...

Страница 22: ...Notes...

Страница 23: ...Notes...

Страница 24: ...353 1 810 7085 ireland invacare com www invacare ie United Kingdom Invacare Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ Tel 44 0 1656 776 222 Fax 44 0 1656 776 220 uk invacare com www in...

Отзывы: