Visual inspection 

Inspect lift unit functions regularly. Check to ensure that material is free from damage. 

Before use

Ensure that the product is correctly assembled.
Check slingbar connection and safety latch function.
Check lift and base-width movement.
Check to ensure that the actuator is correctly installed.

Always read the user manuals for all assistive devices used during a transfer.
Keep the user manual where it is accessible to users of the product.

The lift unit may only be used by persons who have received instruction in the operation of the lift unit.

Safety information

Manual nr: 883-En   Ver. 7   151210    

Service manual - English

SWL: 200 kg/441 lbs

Always read the user manual

Valid from serial number:

MiniLift200

System

RoMedic

TM

200005000

Содержание SystemRoMedic MiniLift200

Страница 1: ...the actuator is correctly installed Always read the user manuals for all assistive devices used during a transfer Keep the user manual where it is accessible to users of the product The lift unit may...

Страница 2: ...e shooting 4 Exploded view and components 5 9 Spare part list 10 11 Changing of spare parts 12 13 Periodic inspection protocol 14 Description for periodic inspection 15 Periodic inspection instruction...

Страница 3: ...button for base width adjustment to widen the base fully and then press the other button to narrow the base again Test lift function by lifting a person not a user using an approved sling At the same...

Страница 4: ...leave the user unattended during a transfer situation Under no circumstances may max load be exceeded See section on Safe working load Ensure that the user s feet do not get caught between the footpla...

Страница 5: ...Exploded view and components 5 SystemRoMedic TM...

Страница 6: ...0 00 Actuator LA28 ML200 1 13 90004033 00 Screw_M6S_DIN933_8 8_M10x35_fzb 2 14 90000539 02 Lock nut DIN 985 CL6 M10 FZB 2 15 90002740 00A Threaded steel rod DIN976 1 8 8 M10x200 fzb 2 16 90000873 01 H...

Страница 7: ...O REV NAME DATE CREATED SHEET 1 OF 1 WEIGHT STATUS TITLE MATERIAL SIZE Released Handwheel with threaded bar ML200 90004012 01 0 03 g 2015 06 10 A4 This drawing and any information or descriptive matte...

Страница 8: ...MATERIAL SPECIFICATION A3 2015 06 10 3 67 g 90004029 Left leg mounted ML200 Released SIZE MATERIAL TITLE STATUS WEIGHT SHEET 1 OF 1 CREATED DATE NAME REV DWG NO 1 5 5 6 E hc makn 2 1 3 4 5 6 Brako ass...

Страница 9: ...cov 10 90004078 00 Screw_RXS_DIN 11 90004079 00 Screw_RXS_DIN 12 90004080 00 Kneep 13 90004012 01 Handwheel w M 14 90002799 00A Screw P6SS DI 15 90003318 00 Nut co 16 90002753 01 Rubb 3 2 1 5 4 6 7 SC...

Страница 10: ...1 set 1 x Rear castor 1 x Locking nut M10 80100074 Bumper kit 1 set 4 x Bumper 80100040 Actuator legspreading LA28 1 set 1 x Actuator LA28 leg spreading 2 x Locking nut 0 25 x Cardboard 80100041 Actua...

Страница 11: ...00 1 pce 1 x Knee pad 80100070 Leg strap plate kit 1 set 2 x Leg strap plate 4 x Screw 80100077 Locking handle knee pad set 1 set 2 x Handle 80100081 Lift arm adjustment pin 1 set 2 x Position bolt 80...

Страница 12: ...at holds the actuator Remove the actuator Remove the rubber cover on the actuator cord and move it to the new actuator Install the new actuator use the accompanying lock nut Connect the actuator to th...

Страница 13: ...d if needed 80100080 Kneepad complete 200 Loosen the locking screw Loosen the lower locking handle and pull the knee pad straight out to remove it Insert the new knee pad in the tube and tighten the l...

Страница 14: ...ging N Product label Test of safe working load Safe working load test upper part Safe working load test base Mechanical emergency lowering Accessories Documentation Instructions manuals Controll box R...

Страница 15: ...Description for periodic inspection 15 SystemRoMedic TM A A G E L J M I K F B A H...

Страница 16: ...ftarm hardware Check the liftarm for visible damage to the surfaces finish etc Check screws nuts bolts etc Kneepad Check the kneepad for visible damage to the surfaces finish etc Make sure it is corre...

Страница 17: ...more space for the weight that is to be lifted NOTE Replace the lower leg support and footrest after test loading Move the lift arm to the outermost position and apply 150 kg Run the actuator 160 mm...

Страница 18: ...nts containing phenol or chlorine as this could damage the aluminium and the plastic materials If disinfection is needed 70 ethanol 45 isopropanol or similar should be used Storage and transportation...

Страница 19: ...The User Manual Medical Device Class I The product complies with the requirements of the Medical Devices Directive 93 42 EEC Lifting speed 35 mm s without load Batteries Two 12V 2 9 Ah valve regulated...

Страница 20: ...each at 600 mm referens height All measurement are in mm inch Tolerance 5 mm 0 2 20 SystemRoMedic TM 860 33 8 425 16 7 410 16 1 295 11 6 960 37 7 600 23 6 1000 39 3 620 24 4 1500 39 880 34 6 50 1 9 75...

Страница 21: ...Reach from base with legs spread to 700 mm referens 21 SystemRoMedic TM 525 20 6 450 875 17 7 34 4 590 1030 23 2 40 5 700 27 5 570 22 4 980 38 8...

Страница 22: ...22 SystemRoMedic TM...

Страница 23: ...23 SystemRoMedic TM...

Страница 24: ...RoMedic offers improvement of both work environment and quality of care and at the same time achieves significant cost savings Always make sure that you have the correct version of the manual The most...