da Vinci SP Instruments and Accessories
Using the Electrosurgical Unit (ESU) with Monopolar and Bipolar Instruments
53
DR
AFT/PRE-RELEASE/C
ONFIDENTIAL
12/5/19
Note: It is recommended to have a backup EndoWrist SP monopolar instrument
available for use to minimize procedure delay in the event of an instrument-related
EnergyShield Monitor fault.
8. Align the arrow on the wide connector end of the monopolar cautery cord with the arrow
above the monopolar receptacle on the EnergyShield Monitor (
) and plug it in.
Ensure that the monopolar cautery cord is fully seated by verifying that the connector
cannot be pushed further into the receptacle. When both ends of the monopolar cautery
cord are properly connected to the EndoWrist SP monopolar instrument and
EnergyShield Monitor, the cord LED on the EnergyShield Monitor illuminates blue (
).
Figure 8.13 All LEDs illuminated blue
9. Confirm all four LEDs illuminate blue (as shown in
) before using the
monopolar instrument.
Figure 8.14 Blue LEDs confirmation
For more information on EndoWrist SP monopolar instruments, see Chapter
starting on page
and Chapter
6 Monopolar Cautery Instrument
starting on
page
.
Cord LED
Arrow above the
monopolar receptacle
Monopolar adapter from
EnergyShield Monitor
Neutral adapter from
EnergyShield Monitor
Patient Neutral Pad
EnergyShield Monopolar
Cautery Cord