da Vinci SP Instruments and Accessories
Using the Electrosurgical Unit (ESU) with Monopolar and Bipolar Instruments
49
DR
AFT/PRE-RELEASE/C
ONFIDENTIAL
12/5/19
EnergyShield Monopolar Cautery Cord
The reusable EnergyShield Monopolar Cautery Cord (PN 430068) is intended to be used in
conjunction with the VIO dV generator, EnergyShield Monitor and EndoWrist SP monopolar
instruments with the appropriate connector (
). The cord is supplied non-sterile and
requires cleaning and sterilization before use. For more information regarding cleaning and
sterilization, refer to the Reprocessing Instructions.
Figure 8.8 EnergyShield monopolar cautery cord connectors
Inspection
Inspect cord for damage or irregularities before and after each use. Visually inspect the cord
for physical damage including:
• Cracked, broken, or otherwise distorted parts
• Broken or significantly bent connector contacts
• Punctures, cuts, nicks, abrasions, unusual bumps, or significant discoloration
• Corrosion of connector contacts
Wide connector
Instrument connector