8

exhalation valve

Under certain circumstances is it necessary to remove the exhalation valve in order to 

clean it. 

1.

  Unscrew the ESA-adapter with a pin spanner or the ESA adapter tool (P/n 

460200022).

Push at the locking tabs on each side of the exhalation valve cover to release it. 

[Fig. 11].

3.

  Remove the exhalation valve assembly [Fig. 12].

.

  Clean and disinfect the exhalation valve assembly according the above 

instructions for the mask. 

.

  Lubricate the pin of the exhalation valve assembly with silicone lubricant 

331900269 and place it back in the hole in the mask connector [Fig. 12]. 

.

  Place the outer end of the exhalation valve spring in the circular cavity on the 

inside of the exhalation valve cover. Now move the exhalation valve cover 

towards the mask and hitch the cover onto the connector by first inserting the 

upper “hook” of the connector into the middle hole (in the upper part) of the valve 

cover. Then push the lower part of the valve cover towards the connector until the 

two locking tabs on both sides lock into position.

BReathinG valve

1.

  Clean the outside of the breathing valve with a cloth and lukewarm water 

(maximum 40° C) with Spirorein or equivalent to remove dirt.

.

  Wipe the breathing valve clean with a cloth and clean water.

3.

  Dry the breathing valve

AttEntiOn!

 WHEN CLEANING THE BREATHING VALVE IT MUST BE COMPLETELY DRY BEFORE IT IS 

REASSEMBLED AND PUT BACK INTO SERVICE. REMAINING WATER MIGHT FORM ICE ON VITAL PARTS, RESULTING 
IN MALFUNCTION OF THE BREATHING VALVE.

extended cleaninG

In certain circumstances it is necessary to disassemble the cover and valve insert in 

order to clean the breathing valve.

1. 

Disconnect the breathing valve.

.

  Unscrew the cover of the breathing valve and remove the positive pressure spring. 

[Fig. 13]

3.

 

Unscrew the locking nut of the breathing valve, use a spanner wrench (tool # 

95992-01) if necessary. Remove the valve insert [Fig. 14], connect a breathing 

hose to pull out the valve insert if necessary.

.

  Clean the cover, locking nut, positive pressure spring and valve housing including 

the diaphragm and diaphragm ring in lukewarm water (maximum 40° C) with 

Spirorein or equivalent. Follow the instructions supplied with the cleaning agent.

.

  Disinfect the diaphragm and valve housing by putting it in a solution with 

Spiroclean, Curacid PSA Ultra or equivalent. Disinfect the lever arm of the valve 

Содержание S-ESA

Страница 1: ...GENAUTOMAT TRENNEN 12 8 REINIGEN UND DESINFIZIEREN 12 9 LAGERUNG 14 10 WARTUNGS UND PR FFRISTEN 14 HANDLEIDING 15 1 VEILIGHEIDSMEDEDELING 15 2 MARKERINGEN AFB 1 16 3 OVERDRUK 16 4 uitvoering van de ge...

Страница 2: ...2 3 4 5 6 1 C B A...

Страница 3: ...7 8 11 10 9 12...

Страница 4: ...13 14 16 15...

Страница 5: ...tective equipment is compatible with the face mask and does not impair the full protection of the respiratory protective device Examples of dangers which may require the use of additional personal pro...

Страница 6: ...ragm Fig 2 The activation arm will automatically return to the starting position after it has been released The button on the diaphragm cover can be pressed to increase the air flow and to activate th...

Страница 7: ...the head harness and loosen the straps 3 Remove the face mask 4 Close the cylinder valve and depressurise the unit by activating the positive pressure 7 DISCONNECT THE BREATHING VALVE 1 Press down the...

Страница 8: ...of the breathing valve with a cloth and lukewarm water maximum 40 C with Spirorein or equivalent to remove dirt 2 Wipe the breathing valve clean with a cloth and clean water 3 Dry the breathing valve...

Страница 9: ...at the lever is correctly located between the diaphragm plate and the roller and at 90 to the roller Fig 15 Screw on the locking nut Fig 16 and tighten moderately 9 Place the positive pressure spring...

Страница 10: ...ist und die Schutzfunktion der Atemschutzaur stung nicht beeintr chtigt Es folgen Beispiele f r Gefahren bei denen zus tzliche pers nliche Schutzvorrichtungen erforderlich sein k nnen Hautsch digende...

Страница 11: ...r Abdeckung f r die Membran gedr ckt wird Abb 2 Der Aktivierungsarm kehrt automatisch in seine Startposition zur ck wenn er losgelassen wird Die Taste auf der Membranabdeckung kann zur Erh hung des Lu...

Страница 12: ...us dem System zu entfernen 7 LUNGENAUTOMAT TRENNEN 1 Dr cken Sie beide Schalter am Lungenautomaten und entfernen Sie ihn aus der Maske Abb 9 2 Lungenautomat vom Mitteldruckschlauch trennen 8 REINIGEN...

Страница 13: ...asser abwischen 3 Lungenautomaten trocknen ACHTUNG NACH DEM REINIGEN MUSS DER LUNGENAUTOMAT VOLLST NDIG TROCKEN SEIN BEVOR ER WIEDER MONTIERT UND IN BETRIEB GENOMMEN WIRD VERBLEIBENDES WASSER KANN AN...

Страница 14: ...r Hebel korrekt zwischen Membranplatte und Rolle sitzt und zwar in einem 90 Winkel zur Rolle Abb 15 Die Sicherungsmutter Abb 16 montieren und leicht anziehen 9 Die berdruckfeder auf die Membran setzen...

Страница 15: ...e bescherming van het ademhalingsbeschermingsapparaat niet verhinderen Voorbeelden van gevaren die het gebruik van aanvullende persoonlijke beschermingsmiddelen kunnen vereisen Vloeistoffen stoom of g...

Страница 16: ...iafragma af te duwen Afb 2 De activeringshefboom gaat automatisch terug naar de beginpositie nadat deze wordt losgelaten De knop op het deksel van het diafragma kan ingedrukt worden om de luchtstroom...

Страница 17: ...gelaatsmasker af 4 Sluit de flesafsluiter en maak de unit drukloos door overdruk in te schakelen 7 DE ADEMHALINGSAUTOMAAT ONTKOPPELEN 1 Druk op de twee knoppen op de ademhalingsautomaat en verwijder...

Страница 18: ...AT 1 Reinig de buitenzijde van de ademhalingsautomaat met een doek en lauwwarm water maximaal 40 C met Spirorein of een overeenkomend product om vuil te verwijderen 2 Veeg de ademhalingsautomaat schoo...

Страница 19: ...rmen op belangrijke onderdelen waardoor de ademhalingsautomaat niet meer goed werkt 8 Monteer het ventielbinnenwerk en zorg dat de hefboom precies tussen de membraanplaat en de rol zit en een hoek van...

Страница 20: ...ansiktsmasken utan att p verka skyddsfaktorn f r andningsskyddet Exempel p faror som kan kr va anv ndning av ytterligare personlig skyddsutrustning V tskor nga eller gaser som kan skada huden Giftiga...

Страница 21: ...ingsventilens vred fr n k pan f r membranet fig 2 Vredet kommer automatiskt terg till utg ngspositionen n r den frig rs Knappen p membranh ljet kan tryckas ned f r att ka luftfl det och aktivera s ker...

Страница 22: ...loss ANDNINGSVENTILEN 1 Tryck in de tv knapparna p andningsventilen och lossa andningsventilen fr n maskanslutningen 2 Lossa andningsventilen fr n andningsslangen 8 RENG RING OCH DESINFICERING Reng r...

Страница 23: ...FELFUNKTION SOM F LJD Ut kad reng ring Vid s rskilda omst ndigheter r det n dv ndigt att ta is r ventillocket och ventilinsatsen f r att kunna reng ra andningsventilen 1 Lossa andningsventilen 2 Lossa...

Страница 24: ...terstra e 47 CH 4133 PRATTELN SWITZERLAND TEL 41 61 827 99 77 FAX 41 61 827 99 70 E MAIL info interspiro ch THE NETHERLANDS BELGIUM INTERSPIRO BV www interspiro nl Operetteweg 35 NL 1323 VK ALMERE NET...

Отзывы: