background image

2.1.5 Factor Selector

In the center of the panel is a bracketed area labeled

FACTOR SHIFT

, which has a button and a one digit

readout. If the

DATA DISPLAY

mode is set to factor,

then the user can select any of 10 different user
programmable sensitivity factors (0-9) that could apply to
different detectors or to different combinations of detector,
filter, and diffuser. These factors are obtained from the
calibration certificate for your particular system. If you
want to read out in current (amperes), dial in 1.000e

0

. See

the next section for changing factors

2.1.6 Changing Factors

There are three BLACK buttons on the left side of the

front panel used for changing the sensitivity factors. They
are marked

‘MSD’

,

‘LSD’

, and

‘EXP’

, which stand for Most

Significant Digits, Least Significant Digits, and exponent,
respectively. The

‘MSD’

button increments the left two

digits of the mantissa, from 1.0 to 9. The

‘LSD’

button

increments the right two digits of the mantissa, from 00 to
99. By holding these buttons down (when in

FACT

OR’

mode), the mantissa digits will increment exponentially up
to a higher value, from 1.000 to 9.99. The

‘EXP’

button

performs a similar function, incrementing the exponent digit
from e-19 to e+19. A full explanation on factor entry is
covered in section 3.2.3.

2.1.7 Bias Voltage

The white button (low center) labeled

5 V

BIAS’,

reverse biases semiconducto r detectors with 5 volts, or it
adds an additional 5 Volt reverse voltage to any vacuum
photodiode for a total reverse selection of 14 volts. In other
words, a silicon detector can be operated without any bias
(photovoltaic mode) for low level and D.C. readings, or with
a 5 volt bias, which increases the speed of response for flash
measurements. A vacuum photodiode always needs a bias,
so it is wired in the connector to give a bias choice of either
9 volts (5 V bias light off), or 14 volts (5 V bias light on).
The bias ON is always the right choice for vacuum
photodiodes (see details in section 3.4.1).

2.1.8 Auto Range

The white button on the lower right side of the front

panel, labeled

‘AU

TO

RANGE’

, gives the user the option to

limit the ranging ability of the system, so one can notice
gross changes in readings at a glance due to either the

‘HI’

indication for too bright, or a very small mantissa for a level
too low. The alternate action of this button selects or de-
selects

‘AU

TO

RANGE’

. If you are over-ranged and wish to

re-establish a new level, you must toggle the function into

‘AU

TO

RANGE’

, then turn off the

‘AU

TO RANGE

mode

again. See section 3.2.2 for more details on the use of this
control.

8

2.1.9 Percentage Mode

The other white button in the lower right corner of the

panel selects a relative mode by making the display reading
equal 100.0, and referencing all subsequent readings to this
original value. These readings are a percentage of that
original value. This is very useful for making transmission
measurement s or reflectance measurements directly. You
must establish a reference for the 100% value, by removing
the filter between the light source and the detector or by
using a white reflectance standard for the 100% condition.

There are many applications that require relative

measurements, such as attenuation measurements through an
optical system, or comparisons between two light sources.
For example you could turn on one light and set to 100%.
Then turn that light off and turn the unknown light on to
read the relative value compared to the first lamp. Since
percentage mode disables the auto-range, the largest relative
reading possible is 1999% and the smallest is 0.1%.

2.2 REAR PANEL

2.2.1 Battery Replacement

In the very upper left corner is a printed message which

gives a brief description of how to insert or replace the
internal batteries. The back panel is tricky to remove,
especially on new units, because everything fits tightly. You
must remove the three nickel-plated Phillips screws on older
models.  Note: The larger screw goes to the right of the
power input connector. On Newer model there are 4 nuts, 2
beside the sensor input and two beside the RS232 connector
and one Philips head screw at the power cord connector.
Once you remove all screws,  yo u must pull the back panel
to the rear by pulling on battery access pull knob (see 6 in
above diagram) with a pair of pliers. To avoid marring the
knob, we suggest you put tape on the knob before you grasp
it with the pliers. While pulling, you should direct your
force up and down to work the assembly to the rear. Be
careful when it lets go, so you do not allow it to move too
rapidly. There are two battery wires that could be broken if
care is not used. Also be careful to notice if one of the
circuit boards is pulled back with the rear panel. It should
NOT be pulled back. Free the board from the rear panel and
carefully push the board back into position so it is even with
the one above it. Set the rear panel assembly on a table
close to the unit without stretching the two wires. You will
now have free access to the battery compartment which
holds six (6)

‘C’

 (NICAD ONLY) cells.

2.2.2 AC Voltage

The ILT1700 will come with the AC switch set to

115VAC, it is very important that you check the switch prior
to use. The two choices are 115 (90-130) Volts AC, which is
common throughout most of North America, and 230 (180-
240) Volts AC, which is common throughou t most of Europe.
In some parts of the world (Japan, for instance), the common
voltage is 100 VAC, which is accommodated on the 115 VAC
selection. You make a selection by inserting a pointed tool,
such as a screwdriver, into the slot, to slide the switch left or
right (115 or 230 volts respectively).

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Содержание ILT1700

Страница 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Страница 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Страница 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Страница 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Страница 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Страница 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Страница 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Страница 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Страница 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Страница 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Страница 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Страница 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Страница 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Страница 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Страница 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Страница 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Страница 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Страница 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Страница 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Страница 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Страница 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Страница 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Страница 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Страница 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Страница 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Страница 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Страница 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Страница 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Страница 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Страница 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Страница 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Страница 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Страница 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Страница 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Страница 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Отзывы: