1.1.1 International Power Settings

Switch AC voltage selector switch to 220 or 115 VAC as

applicable. Select 115 VAC for use on power systems
between 90 and 130 VAC (JP, US, CA). Select 230 VAC for
use on power systems between 180 and 260 VAC (EC, AU).

1.1.2 Power Cord

Connect power cord to power input then into main wall

socket. The standard male power plug on the back of the
ILT1700 accepts power cords from many different countries
and power systems.

1.1.3  Select AC Power Source

Switch AC / BAT selector to AC position for operation

using an alternating current power supply. The internal
lithium batteries will automatically recharge whenever the
ILT1700 is plugged into an external power supply. The
rechargeable Batteries(NICAD ONLY) must initially be
charged before the first use on battery power.

1.1.4 Plug In Detector

Connect detector to sensor input via the 15 pin D-Sub-

Min connector labeled SENSOR INPUT. The D connector
is wider on the top than on the bottom. The plug must be
aligned properly with the socket for the two to mate.

1.1.5 For More Information

Refer to section 2.2 for more detailed information

1.2 Front Panel Set Up:

1.2.1 Turn Power On

Press POWER switch. AC LED indicator should be

illuminated. If not check rear panel switch to ensure it is in
the AC position. (see section 10.4 if unit does not power up)

1.2.2 Store Sensitivity Factors

Enter sensitivity factor from detector calibration

certificate.

a. Press Factor Display button.
b. Press FACTOR SELECT button until 0 appears in
window above button.
c. Enter sensitivity factor by pressing MSD,LSD &
EXP buttons. Enter the number and exponent as it is
shownon your calibration certificate. This factor is now
permanently stored as factor number 0.
d. Repeat steps b & c for additional factors. Ten factors
can be stored in this manner. Remember to increment
button to next factor number before entering new factors.

1.2.3 Display Data

Press DATA/FACTOR/DISPLAY button. ILT1700 is now
ready to display data.

1.2.4 Bias Selection

Press 5V Bias button if you are using vacuum photodiode
type detector (SED240 or SED220), or for flash
measurements.

1.2.5 Zero The Detector

Cover detector with opaque object, wait ten seconds and
press ZERO button. For ambient zero simpl y press zero button
for over one second and release. (will subtract current zero from
all future readings) To return to internal zero, unplug the
detector and press zero.

1.2.6 Select Measurement Mode

Select measurement function by pressing either D.C. (signal)
button or INT button for continuous or integrated readings.
Remember INT will sum up the signal over that integration
time and D.C. mode will read the average steady state
light levels.

1.3 Begin to Measure

1.3.1 Install Detector

      Plug a detector into the Sensor Input. Select the
appropriate calibration factor for that sensor combination
using the factor select button.

1.3.2 Zero the Detector

     First you must decide if you want a dark zero, or an
ambient zero. An ambient zero will subtract the ambient light
levels  from future readings, remove the black protective cap
and press zero for over one second. If you want a dark zero,
assure the black cap is on or cover sensor with an opaque
object, wait 10 seconds then press zero for over  one second.
Wait for zero LED to turn off. (for high gain and thermopile
detectors, do not hold the sensor while zeroing , and be sure to
completely block all light with an opaque object )

1.3.3 Operating in D.C. Signal Mode

The ILT1700 default operating mode is D.C.(signal)

mode. If the D.C. LED is not lit, press the D.C. button to
enter D.C. mode. The ILT1700 samples continuously,
updating the display every half second.

To measure as a percentage of a known peak value, first

establish the peak condition, then press the SET 100% button.
ILT1700 reads as a PERCENTAGE of the set value.
      To measure within a fixed range, turn off the Auto Range
feature by pressing the AUTO RANGE button. The exponent
on the display will not change when the Auto Range LED is
off. The ILT1700 default setting is Auto Range On.

Hold a reading temporarily by pressing the

HOLD button. This freezes the display for use in a
darkroom.

1.3.4 Operating in Integrate Mode

Press the INT button to begin integrating

(summing) all readings. This feature is useful for making
dose measurements.

Press the HOLD button to temporarily pause

the display. The integration continues, however, allowing
you to write down interim readings.

Press the HOLD button again to continue

displaying the ongoing integration.

Note: the Zero level that you set previously is

subtracted continuously from all integrated readings. You
can reset the Zero and integral in Integrate mode by pressing
the ZERO
button in the dark or ambient condition.

Press the D.C. button to stop the integration.

6

1.1 Power

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Содержание ILT1700

Страница 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Страница 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Страница 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Страница 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Страница 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Страница 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Страница 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Страница 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Страница 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Страница 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Страница 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Страница 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Страница 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Страница 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Страница 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Страница 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Страница 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Страница 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Страница 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Страница 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Страница 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Страница 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Страница 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Страница 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Страница 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Страница 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Страница 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Страница 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Страница 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Страница 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Страница 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Страница 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Страница 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Страница 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Страница 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Отзывы: