International Light Technologies ILT1700 Скачать руководство пользователя страница 14

3.3.1 Integration Zero

A

‘ZERO’

state for integration is when the first

derivative of the output is zero. In other words, a non-
changing output . To get this state, the integral of any
unwanted input signal must be measured and subtracted from
the integral of the total signal.

‘Unwanted ’

can mean ambient

light or some detector dark current contribution. An example
may be the best way to shed some

‘light’

on the subject.

Suppose we are in a room with two table lamps in it, called

A

and

‘B’.  ‘

A

is left on so you can see what you are doing,

and

‘B’

is the sample lamp to be measured. With

‘B’

off, you

would press

‘ZERO’

so the signal dose from

A

will be

ignored.  The ILT1700 must measure this incoming signal
from

A

’, whic

h will subsequently be removed from the total

integral. This requires subtracting the product of

A

times the

integral time from the total integral produced by

A

and

‘B’

together. Fortunately, this is all done for you by the internal
computer. Another thing that is done automatically is the
zeroing of the dose accumulation registers at the time you
press the

‘INT’

button. If you zeroed on

A

and never

turned

‘B’

on, the accumulated dose should be essentially

zero, or equal to any small change in

A

as compared to its

value when the

‘ZERO’

button was pressed. This change

can be positive or negative, depending upon whether

A

got

a little brighter or dimmer, respectively. It is reasonable to
make sure your sample lamps dose is much greater than
any zeroed dose, to be sure that small changes in the
ambient will not effect the accuracy of the final result. A
good rule of thumb is to try to keep your signal
contribution at least 10 times larger than the ambient
contribution. 
This may require workin g in the dark,
which can be done by either remote control from another
room or by pressing the hold button while still in the dark,
and then taking the data with the lights turned on (See
section 3.3.6 for more on this).

3.3.2 Long Term Integration

Any integration longer than one second is considered

long term integration. In this state, the ILT1700 keeps up
with the incoming signals, and presents an updated readout
of the progress every 1/2 second. A proper zero should be
done prior to use of integration.

As soon as the

‘INT’

button is pressed, the integration begins. When the

‘HOLD’

button is pressed the integration ends and reads out the final
integral. If you wish to continue integration without
restarting, adding the new dose to the old dose, just press

‘INT’

again,

and ‘HOLD’

when you want to stop. If you

want to start over again from zero, then press

‘D.C.’

or

‘ZERO’

before

the ‘INT’

button. If you press

‘INT’

two

times in a row, the system will start a new integration at the
time of the second press. This can be useful for multiple
exposures, assuming the light is momentarily off to permit
time for readout or transfer to a computer or printer. A press
of

‘INT’

starts the next cycle without any other buttons

being pressed.
     There is no limit to how long you can integrate. The
system will maintain perfect accuracy for over 18 years even
if the signal is the maximum permitted into the ILT1700.
There is no loss of accuracy for extended integrations, and
the auto ranging keeps track of the magnitude no matter how
large it gets.

 3.3.3 Flash measurements

The procedure for making flash measurements requires

the use of three other function buttons. We start with the

‘5

V BIAS

button. IT IS NECESSARY to reverse bias the

detector to make the detector respond faster, as well as to be
able to handle the high peak currents without saturating and
also to provide bias compliance to an input charge circuit
which temporarily holds the charge long enough for the
precision charge measurement to be performed. In other
words the red

‘5

V BIAS

LED should be lit when making

any flash measurement at speeds faster than one second.
The

‘ZERO’

button is used to cancel out any unwanted

ambient contributio n that is present before and during the
flash (see previous section). Next, press the

‘INT’

button

just before the flash is to occur, or if you wish, you may use
the

‘FLASH not ’

output line (pin 9) to activate the flash

event (see section 6.1 for details). The summation registers
will be instantly zeroed and the integration will commence.
Two and a half (2.5) seconds after the flash, press the

‘HOLD’

button to freeze the final result. This extra time is

required to fully dump the charge from a temporary holding
circuit on the front end of the input system and be displayed
by the computer. This permits high speed measurements to
occur faster than the computer can keep up with the input.
This feature allows the system to handle 6 decades of high
speed charge ranges at a sacrifice of summation speed. In
fact, it can take up to 2.5 seconds to produce a final answer
to 0.1% resolution. This does not pose any difficulty in any
application that can operate in

a ‘one

-shot

mode. However,

if there are multiple flashes, you must use a shutter to
capture one flash, or integrate for a known number of
flashes, and divide the answer by that number. Often it is
possible to turn off the power supply to the flash lamp after
the required number of flashes has occurred. This will
permit the summation time (2.5 seconds) to occur.

3.3.4 Averaging

As mentioned above on the subject of multiple flashes,

the system can be used to find the average of many flashes.
This is done by integrating throughout many flashes. By
maintaining a count of the flashes and dividing by that
number, you will arrive at the average output from one flash.
Another form of averaging is that which occurs as the result
of a very irregular light output.  An example of such a
source is an arc welding process. If you wished to find the
average ultraviolet dose the welder might receive during a
work day, you could leave the Radiometer running in the
integrate mode all day long, or you could integrate for a
reasonable length of time to determine the dose for part of a
day, then multiply by the number of daily portions
representative in the work day to find the total accumulation

13

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Содержание ILT1700

Страница 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Страница 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Страница 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Страница 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Страница 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Страница 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Страница 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Страница 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Страница 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Страница 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Страница 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Страница 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Страница 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Страница 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Страница 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Страница 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Страница 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Страница 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Страница 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Страница 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Страница 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Страница 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Страница 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Страница 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Страница 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Страница 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Страница 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Страница 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Страница 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Страница 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Страница 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Страница 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Страница 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Страница 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Страница 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Отзывы: