17

4.5.4  Software

The  ILT1700.VI  driver  is  a  sample  LabVIEW

®

  program

that  allows  the  end  user to  perform  data  acquisition  with  an
ILT1700  Research  Radiometer.    The  VI  will  run  in
LabVIEW

®

  8.2  Full  Development  System  and  offers  the

ability  to  record  readings  in  either  DC,  or  integrate  mode  as
well as to calculate the percentage of a baseline reading .  The
data  is  saved  in  a  tab -delimited  text  file  that  can  be  opened
with  Excel.    Users  with  LabVIEW

®

  8.2  Full  Development

System can customize the VI to meet their unique application
requirements.

We also offer an ILT1700.VI program for customers who

do not own a copy of LabVIEW

®

 Full Development System.

The application is a fully compiled, run -time version of the VI
driver.  Users can run the program, obtain data from an
ILT1700 meter and save it into a file.  The program is
available for Windows 2000, XP and VISTA.

After you install your COM PORT drivers (4.5.1) you will

need to install the complim entary Virtual ILT1700-USB
program. (Full installation and operation instructions are
provided in the file named ILT1700_README on the CD.)
This software program provides the maximum flexibility and
functionality available without sacrificing ease of use. This VI
executable displays incoming data in real time and allows the
user to save data to a text file for importation into a
spreadsheet program.

4.3

NEW

USB and RS232

The current model ILT1700 now has improved features for

computer connection. T he first is the inclusion of a new USB
2.0 compatible (backwards compatible with USB 1.1) output
port for sending measurement data to any USB equipped PC.
The second upgrade is the replacement of the older 25 -pin D-
sub, RS232 serial output connector with a  more commonly
available 9-pin D-sub connector.  Note: The 0-1V output jack
has been eliminated on the rear panel,  but 0-1V output is still
available using the accessory I/O connector, refer to section
6.5 for further information)

4.3.1 Setting up a Com port

If you intend to use the new USB port please note the

following: You must install a USB driver prior to using the
USB port.  You should have received a CD with your
purchase.  If you do not have the CD you will need to contact
ILT to obtain a copy.

Upon plugging the ILT1700 into a USB port on a PC, the

computer’s operating system will detect the instrument and

will need to be pointed to the directory where you have
extracted the driver .zip file in order to properly install the
driver.  The installer will install a virtual Com port for the new
device. Please make a note of which com port you have
selected as you will need thi s information when running the

LabVIEW®

 drivers. _____________

4.5.2 Serial output

Serial output information for both  the RS232 and USB port is

as follows:

Baud rate:     4800

Start bits:

1

Data bits:      8

Stop bits:

1

Parity:

       None

Hardware:     None or Hardware

4.5.3 PC cable connection

IL1700 to PC cable:  DCE to DCE null modem male to
female:

IL1700 end (DB9 Male)

PC Cable End (DB9 Female)

Pin 3 Transmit data-------------------------Pin 2 Receive data
Pin 5 System Ground-----------------------Pin 5 System ground
Pin 7 RTS------------------------------------ Pin 8 CTS
Pin 4  DTR ----------------------------------Pin 6 DSR

代理美国International Light辐照计http://www.testeb.com/yiqi/ilt/zhaoduji.html 深圳市格信达科技 电话18823303057 QQ:2104028976

Содержание ILT1700

Страница 1: ...ILT1700 INSTRUCTION MANUAL Revised 1107 jmf International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Страница 2: ...means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests should be made through t...

Страница 3: ...nt Panel 7 2 1 1 Power Controls 7 2 1 2 Function Controls 7 2 1 3 Readout Display 7 2 1 4 Display Selector 7 2 1 5 Factor Selector 8 2 1 6 Changing Factors 8 2 1 7 Bias Voltage 8 2 1 8 Auto Range 8 2...

Страница 4: ...put 15 4 2 1 Voltage Range 15 4 2 2 Auto Ranging Considerations 15 4 2 3 Negative Readings 15 4 2 4 Output Impedance 16 4 2 5 Character Format 16 4 2 6 Word String Format 16 4 3 NEW USB AND RS232 170...

Страница 5: ...Cycle fast pulse 26 8 10 2 High Peak Amplitude 26 9 General Specifications 27 9 1 Current Measurement Accuracy 27 9 2 Optical Accuracy 27 9 3 Radio Frequency Interference 27 9 4 Size and Weight 27 9...

Страница 6: ...S SE EL LE EC CT T S SW WI IT TC CH H P PR RI IO OR R T TO O T TU UR RN NI IN NG G O ON N Y YO OU UR R I IL LT T1 17 70 00 0 NOTE The Accessory input ouput connector may be marked ACCESSORY INTPUT Ho...

Страница 7: ...ure readings To return to internal zero unplug the detector and press zero 1 2 6 Select Measurement Mode Select measurement function by pressing either D C signal button or INT button for continuous o...

Страница 8: ...essing INT the unit begins to sum up all the energy over time until you press HOLD which terminates the integration and displays the final results Integration is also useful for flash integrations and...

Страница 9: ...readings to this original value These readings are a percentage of that original value This is very useful for making transmission measurements or reflectance measurements directly You must establish...

Страница 10: ...d SENSOR INPUT This of course is where you plug in the light sensor or detector for measuring another parameter The correct internal voltages for all International Light detectors are provided in the...

Страница 11: ...ements a carry over from the electronics field meaning Direct Current or signal mode In other words we are referring to those measurements where the magnitude of the light level remains reasonably sta...

Страница 12: ...you wish to change 0 to 9 Now use the three black buttons marked MSD LSD AND EXP to enter the number in scientific notation just as it appears on the calibration certificate By holding a button in th...

Страница 13: ...ange in light magnitude and the tremendous variation in measurement units As an example your eye can see in an environment which can have a brightness change of one million to one If you couple this w...

Страница 14: ...integrate The system will maintain perfect accuracy for over 18 years even if the signal is the maximum permitted into the ILT1700 There is no loss of accuracy for extended integrations and the auto r...

Страница 15: ...n D C readings between 9 and 15 volts bias is only about 1 but to gain that extra accuracy use it with the 5V BIAS light on 14 3 4 2 Flash Bias Detector bias has the most significant effect on system...

Страница 16: ...ange The voltage is designed to be compatible with most strip chart and X Y recorders and yet provide a large enough signal to avoid excessive noise pickup For these reasons we have selected a range f...

Страница 17: ...r the displayed data to other equipment such as a printer computer or modem There are four different modes of readout auto range fixed range percent and factor In the first two modes the interface sen...

Страница 18: ...SB 2 0 compatible backwards compatible with USB 1 1 output port for sending measurement data to any USB equipped PC The second upgrade is the replacement of the older 25 pin D sub RS232 serial output...

Страница 19: ...For very flat response and long wave detection we offer the SED623 thermopile detector which has a built in preamplifier to transform the tiny light induced voltage signal into an amplified current co...

Страница 20: ...out any ambient conditions Some time later when a test lamp was in position INTG would be pulled low for more than 50 milliseconds or until FLASH was sensed Three seconds after that HOLD would be pull...

Страница 21: ...r than two digit resolution from 1e 12 to 2e 3 Amperes and read conductance from 2e 13 to 4e 4 Siemens formerly mhos each of which cover more than nine 9 decades In addition it has the ability to inte...

Страница 22: ...egral of the Effective 8 2 FLUX DENSITY MEASUREMENTS Most light measurements will be of this type where flux refers to the rays of light and density refers to the number of rays per unit area falling...

Страница 23: ...lications that either sense the light directly on the surface of the cell or on the cell behind a transmissive filter the reference distance is slightly in front of the cell surface Each piece of glas...

Страница 24: ...ance squared In other words if you were 10 centimeters away from the rim of the port you would divide 11 10 by 100 and get the solid angle to be 0 111 sr If the sphere is calibrated to read optical wa...

Страница 25: ...ement of only the line of sight rays Be careful that the detector is looking at an area located in the uniform part of the test surface If you back away from the surface too far the input angle will e...

Страница 26: ...avelength 8 9 1 Lambertian Response Lambertian response is in reference to a particular angular response proportional to the cosine In other words the cosine of the angle normal to the face of the det...

Страница 27: ...and in a capacitor on the front end of the system By measuring with a 5 volt reverse bias on the detector we now eliminate junction saturation due to internal series resistance voltage drop in the det...

Страница 28: ...the computer board then insert the amplifier board and reinsert the computer board in last If the readings do not change with the input you probably have a problem with the amplifier board or the det...

Страница 29: ...o plugging the ILT1700 into 230VAC Customer should always double check the power and source selections on the rear panel prior to plugging in the ILT1700 To change th e fuse follow the steps listed in...

Страница 30: ...ER D SPARE E SPARE F NO CONNECTION H NO CONNECTION J INTG INTEGRATE K HOLD L SWDN SWITCH DOWN M ACCS ACESSORY N GROUND RS232 9 PIN CONNECTOR 1 GROUND PROTECTIVE GROUND 2 TRD TRANSMIT DATA NOT CONNECTE...

Страница 31: ...29 International Light http www testeb com yiqi ilt zhaoduji html 18823303057 QQ 2104028976...

Страница 32: ...INT LAST VALUE STORED 260 LOCATE 19 53 PRINT 270 LOCATE 19 50 PRINT A X 280 LOCATE 17 71 PRINT MEMORY 290 LOCATE 18 70 PRINT LOCATION 300 LOCATE 19 72 PRINT X 310 SELECT INPUT 1 320 LOCATE 11 1 PRINT...

Страница 33: ...RT DRIVER VCP You will need your ILT1700 meter a USB cable and your complimentary drivers CD Connect the ILT1700 with a USB cable to a USB 2 0 port of your computer The Found New Hardware Wizard will...

Страница 34: ...using this information can be found in the Device Manager of your computer Use Control Panel System Hardware Device Manager Ports in the Start Menu for information The default COM port is set at COM4...

Страница 35: ...k the Set 100 button to set the baseline 100 level All future readings will be in percentage as a comparison to the 100 baseline reading To measure accumulated dosage or total exposure press the Integ...

Отзывы: