Technical Data
Accuracy Class
Analog
0.1%
Digital
0.05%
Inputs
Measurement Range
±1mV/V
(JP1 on 1 with 5V sensor supply)
±2mV/V or ±3.5 per software)
±3.5mv/V per software)
(JP1 on 2 with 2.5V sensor supply)
Connectible full bridge
4 x 350 Ohm
Bridge supply voltage
2.5V/5V
Input Impedance
>20MOhm (300pF)
Common mode rejection
DC
>120 dB
100Hz
>100 dB
Analog Input 1
Input voltage range
0 - 10V
Input resistance
56 kOhm
Accuracy
Linearity deviation
<0.02% of unit
Influence of temperature on the zero
point per 10K
Measurement range 1mV/V: <0.4 type 0.2% of unit
Measurement range 2mV/V: <0.2 type 0.1% of unit
Influence of temperature on the measurement sensitivity per 10K referring to the
measured value
Analog output
<0.1; type 0.05%
Display / digital
<0.1; type 0.005%
Resolution
Peak Value
>30000 parts
RMS
>150000 parts
Output
Analog Output Filter
-3dB cut-off frequency, Bessel,
2. order
3.5; 260; 1700 Hz (can be switched using software)
Digital Output Filter
3dB cut-off frequency
0.06-1700 Hz
Digital output measuring
rate
0-3750 Hz
Analog Output
±5V
Source resistance
47 Ohm