Measuring Amplifier Menu
The measuring amplifier menu is used to set the parameters of the measuring amplifier, for example:
Sensor scaling factor
Unit
User-definable offset
Parameter memory
The measuring amplifier main menu can be reached by pressing the MENU button from the
measured value display.
Menu Entry Level 1 Menu Entry Level 2
Menu Entry Level 3
Sensor config.
➔
unit
Select unit
➔
Sensor capacity
Numeric setting of the physical nominal value of the
sensor
➔
Rated output
Numeric setting of the electrical characteristic value
of the sensor
Strain analysis
➔
Set gage factor
➔
Numeric setting of the K-factor between 0.2 and
2583
➔
Set bridge type
➔
Full bridge: Full bridge circuit with 4 individual
DMS, all in longitudinal direction
➔
Half bridge: Half bridge circuit with 2 individual
DMS, both in longitudinal direction
➔
Quarter bridge: Quarter bridge circuit with one
DMS
➔
PR.full bridge: Full bridge circuit with 4 individual
DMS, 2 in longitudinal direction and 2 in cross
direction
➔
PR.half bridge: Half bridge circuit with 2 individual
DMS, one in longitudinal direction and
one in cross direction
Load settings
➔
default:
Manufacturer’s
settings i.e. restoring
GSV-2 parameters to
default settings
➔
user 1:
Userconfigurable
data set no.1 i.e.
loading the
parameters that were