9330 Operating Manual
Interface Inc.
● 7401 East Butherus Drive, Scottsdale, Arizona 85260 USA ● Phone 480.948.5555 ● Fax 480.948.1924
www.interfaceforce.com
● Email:
●
800.947.5598
Document Number 15-216 Rev B
Page 15 of 74
Logger Menu
The logger menu has the following functions:
View and set data recording mode on the SD memory card,
Select USB connection mode,
Select display illumination characteristics,
Display battery life,
Set date and time,
Select additional functions e.g. display and recording of maximum, minimum and average
measured values.
The Logger menu can be reached by pressing the MODE button from the measured value
display or via the last entry of the measuring amplifier main menu.
Menu Entry Level 1
Menu Entry Level 2
Menu Entry Level 3
Logging ON, OFF,
onOK
➔
Set Log: ON
– Permanent recording of
measured data on the SD card
➔
Set Log: OFF No Recording
Set Log: onOK - Recording of measured data on
the SD
card as long as the OK button (at measured value
displayed) is pressed
USBmode: Comm,
SDcard, none
➔
USBmode: SDcard The device is a Mass
Storage
Device when connected to the USB port, which
provides
access to the data on the SD card. Simultaneous
recording of measured value to file is not possible.
If
this mode is activated, recording is switched off.
➔
USBmode: Comm The device is in serial USB
mode.
Our communication programs (e.g. GSVcontrol,
GSVmulti) can then be used. A standard driver is
loaded
when connected for the first time.
➔
USBmode: none USB is switched off.
Bat: level bar or
Charge (with
Percentage display)
➔
USB power: ON, OFF
If power supply is via USB port is enabled, the
GSV-
2MSD-DI can be supplied by the USB bus if it is
connected to a PC and configured as an USB
device. If the device is switched off, the battery
can be charged.
➔
Set: USBpwr ON
➔
Set USBpwr OFF
Switch power supply