Portable and mobile RF communications equipment can affect the Eclipse. Install and operate the Eclipse according to the EMC information
presented in this chapter.
The Eclipse has been tested for EMC emissions and immunity as a standalone instrument. Do not use the Eclipse adjacent to or stacked with
other electronic equipment. If adjacent or stacked use is necessary, the user should verify normal operation in the configuration.
The use of accessories, transducers and cables other than those specified, with the exception of servicing parts sold by Interacoustics as
replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the device.
Anyone connecting additional equipment is responsible for making sure the system complies with the IEC 60601-1-2 standard.
Guidance and manufacturer’s declaration
- electromagnetic emissions
The Eclipse is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse should
assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Eclipse uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Eclipse is suitable for use in all commercial, industrial, business,
and residential environments.
Harmonic emissions
IEC 61000-3-2
Not Applicable
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not applicable
Recommended separation distances between portable and mobile RF communications equipment and the Eclipse.
The Eclipse is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user
of the Eclipse can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Eclipse as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum output power
of transmitter
[W]
Separation distance according to frequency of transmitter
[m]
150 kHz to 80 MHz
d
= 1.17
√P
80 MHz to 800 MHz
d
= 1.17
√P
800 MHz to 2.5 GHz
d
= 2.23
√P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where
P
is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note 1
At 80 MHz and 800 MHZ, the higher frequency range applies.
Note 2
These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Guidance and Manufacturer’s Declaration
- Electromagnetic Immunity
The Eclipse is intended for use in the electromagnetic environment specified below. The customer or the user of the Eclipse should assure
that it is used in such an environment.
Immunity Test
IEC 60601 Test
level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be greater than 30%.
Electrical fast
transient/burst
IEC61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines
Not applicable
+1 kV for input/output lines
Mains power quality should be that of a
typical commercial or residential
environment.
Surge
IEC 61000-4-5
+1 kV differential mode
+2 kV common mode
Not applicable
Mains power quality should be that of a
typical commercial or residential
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
lines
IEC 61000-4-11
< 5%
U
T
(>95% dip in
U
T)
for 0.5 cycle
40%
U
T
(60% dip in
U
T)
for 5 cycles
70%
U
T
(30% dip in
U
T)
for 25 cycles
<5%
U
T
(>95% dip in
U
T)
for 5 sec
Not applicable
Mains power quality should be that of a
typical commercial or residential
environment. If the user of the Eclipse
requires continued operation during
power mains interruptions, it is
recommended that the Eclipse be
powered from an uninterruptable power
supply or its battery.
Содержание Eclipse ABRIS
Страница 1: ...Instructions for Use EN Eclipse Science made smarter D 0120588 B 2019 03...
Страница 2: ......
Страница 80: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 72...
Страница 112: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 104...
Страница 132: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 124...