9

 

Operating instructions ASSR 

9.1  Using the ASSR module 

 

 

1.  All contact between the conductive parts of electrodes or their connectors, including the neutral 

electrode and other conductive parts including the earth must be avoided. 

 

2.  Please check the setup before recording and verify that the correct stimulus rate and stimulus level 

will be used as another operator/person may have changed/deleted the protocol setting. During 
recording the stimulus rate and level can be seen on the user interface. 
 

3.  If the system has not been used for a while, the operator should inspect the transducers (e.g,. check 

the insert phone silicone tubing for cracks) and electrodes (e.g., check the expiry date of the 
disposable electrodes, check cables for damage) to verify that the system is ready to start testing 
and provide accurate results.  

 

4.  Only electrode gel intended for electroencephalography must be used. Please follow the 

manufacturer’s instructions regarding the use of the gel.

 

 

5.  The presence of tinnitus, hyperacusis or other sensitivity to loud sounds may contraindicate testing 

when high intensity stimuli are used.  

 

 NOTICE

 

1.  The Eclipse system consists of a 2 channel input board which allows the user to create 

measurements for both ears without switching the electrodes.   
 

2.  In the case of tense patient muscles, particularly in the region of the neck, nape and shoulders, the 

quality of the recordings may be poor or completely rejected. It may be necessary to reinstruct the 
patient to relax and then resume testing when the muscles have relaxed.  

 
 
 
 
 
 
 
 
 
 
 

Содержание Eclipse ABRIS

Страница 1: ...Instructions for Use EN Eclipse Science made smarter D 0120588 B 2019 03...

Страница 2: ......

Страница 3: ...5 Driver installation 13 2 4 6 Restoring factory default settings in the software 14 2 4 7 Installing a language pack 14 2 5 Reader station 15 2 6 License 15 2 6 1 Eclipse serial number 15 2 6 2 DSP...

Страница 4: ...l curves comparing to a historic session 33 3 2 10 Merging curves creating a summed curve 33 3 2 11 Adding curves 34 3 2 12 Differential Ipsi minus contra Ipsi Contra curve 34 3 2 13 Differential A mi...

Страница 5: ...ing 55 3 14 Making a P300 MMN recording 56 3 14 1 Example of electrode montage for P300 MMN 56 3 14 2 Available P300 MMN stimuli 56 3 14 3 Summary of parameters for P300 and MMN 57 3 15 System perform...

Страница 6: ...odule 90 7 6 Using the Print Wizard 95 8 Operating instructions ABRIS 97 8 1 Using the ABRIS module 97 8 2 Mounting electrodes 97 8 3 Impedance check 98 8 4 Loop back LBK15 testing 98 8 5 ABRIS test s...

Страница 7: ...6 Extending the test time 114 9 8 7 Adjusting the stimulus intensity 114 9 8 8 Stopping a test frequency intensity 115 9 9 The audiogram tab 116 9 9 1 Estimated audiogram 116 9 9 2 Selected correction...

Страница 8: ...11 8 2 EP15 EP25 VEMP module stimulus maximum intensity 138 11 8 3 TEOAE module 139 11 8 4 DPOAE module 139 11 8 5 ABRIS module 139 11 8 6 ASSR module 140...

Страница 9: ...1 5500 Middelfart Denmark Tel 45 6371 3555 Fax 45 6371 3522 E mail info interacoustics com Web www interacoustics com It is the purpose of this manual to provide users of Interacoustics EP15 EP25 VEM...

Страница 10: ...Emissions The target population for Eclipse TEOAE includes all ages DPOAE The Eclipse DPOAE system is intended for use in the audiologic evaluation and documentation of ear disorders using Distortion...

Страница 11: ...external auditory canal Testing should not be performed on patients with such symptoms without a medical doctor s approval Testing VEMP patients with neck muscular and cervical injury problems should...

Страница 12: ...Electrode Cables with Buttons ETSE tab surface electrode kit Jumper Cable 125mm NuPrep gel 4oz 114g tube SPG15 Gauze Swabs PEG15 Pregel foam snap electrodes 25 pcs 1 Disposable Snap electrodes1 Dispos...

Страница 13: ...aneously 2 A Separation Device isolation device is needed to isolate the equipment located outside the patient environment from the equipment located inside the patient environment In particular such...

Страница 14: ...s connected to the corresponding ear of the patient and that the correct test ear is selected from within the user interface NOTICE 1 To prevent system faults take appropriate precautions to avoid PC...

Страница 15: ...ould be reported immediately to the supplier of the instrument together with the invoice serial number and a detailed report of the problem In the back of this manual you will find a Return Report whe...

Страница 16: ...he CE mark indicates that Interacoustics A S meets the requirements of Annex II of the Medical Device Directive 93 42 EEC T V Product Service Identification No 0123 has approved the quality system Yea...

Страница 17: ...are 2 3 Hardware installation When connecting the Eclipse to the mains supply and to a computer the following warnings must be observed 1 All contact between the conductive parts of electrodes or thei...

Страница 18: ...le The computer can be attached to other devices and still maintain patient safety Please consult with a medical device safety specialist if in doubt The separable mains plug for is used to safely dis...

Страница 19: ...input output E g eABR cochlear implant testing etc 6 Talk Forward Connector for talk forward microphone 7 OAE Plug for the OAE probe Note the OAE probe is dedicated to one Eclipse only Calibration is...

Страница 20: ...acoustics will not make any guarantee to the functionality of the system if third party software is installed with the exception of the OtoAccess or a Noah 4 0 or higher 2 When running the EP15 EP25 V...

Страница 21: ...guided through the installation process Follow the prompts that appear on the dialog boxes 3 During installation you may be asked to install default protocols correction factors normative data report...

Страница 22: ...le click the USB drive to view the contents of the installation USB Double click the setup exe file to initiate the installation The following languages are available on the Eclipse software installat...

Страница 23: ...distributor and inform them of the Eclipse serial number the DSP serial number and your current license key for this instrument 2 6 1 Eclipse serial number The Eclipse hardware serial number is locate...

Страница 24: ...etected To start from OtoAccess 1 Open OtoAccess 2 Select the patient you want to work with by highlighting it blue 3 If the patient is not yet listed press the New client button fill in at least the...

Страница 25: ...ts tab 2 Create a new instrument by a Type in the Software module name in the New Instrument name field b Select the relevant module from the Software modules dropdown menu c Select Eclipse from the H...

Страница 26: ...you want to work with by highlighting it orange 3 If the patient is not yet listed press the Add a New Patient button fill in at required fields save the patient details by pressing the OK button 4 D...

Страница 27: ...st be used Please follow the manufacturer s instructions regarding the use of the gel 5 The presence of tinnitus hyperacusis or other sensitivity to loud sounds may contraindicate testing when high in...

Страница 28: ...t Delete waveform markers on all curves to delete all waveform markers for the selected curve View provides access to waveform display options 1 Select Left to view only the recorded left ear curves o...

Страница 29: ...ect a Test Protocol from the dropdown menu 3 1 5 Temporary setup The Temporary Setup icon allows temporary changes to be made to the selected protocol Changes will only apply to the current test sessi...

Страница 30: ...n displays the contralateral waveform for the selected waveform 3 1 12 Talk forward The Talk Forward icon activates the talk forward function The test will pause when this function is activated The ta...

Страница 31: ...ting a historic session the session date remains unchanged in the database as this always refers to the date of the recording 3 1 16 Save Exit The Save Exit icon saves the current test session and clo...

Страница 32: ...rding more than one intensity can be selected The highest intensity will be presented first Use the Next Intensity button to move to the next intensity without stopping the test 2 Stimulate Stim Stimu...

Страница 33: ...he next intensity selected in the Manual Stimulus window 3 1 26 Fmp residual noise graph The Fmp Residual Noise Graph provide information about the quality of the selected curve 1 The Quality Target e...

Страница 34: ...3 1 29 Advanced EEG Double click on the Raw EEG graph to open the Advanced EEG graph 1 Adjust the rejection level display using the arrows 2 The grey area indicates the area where rejection has been...

Страница 35: ...line appears with the actual timing for hiding the artifact all waveforms in once 3 Point with the mouse at the black line end and the mouse symbol is changing for user edition click and drag the line...

Страница 36: ...keyboard Change the Display Gain for a single curve by right clicking on the selected waveform s handle and selecting the Enlarge or Diminish option Alternatively use Ctrl the up or down arrow key on...

Страница 37: ...36 Adding a note to the waveform Right click on the curve handle and click on the option Note Type here the waveform note comment The note has a maximum of 20 characters to keep space for the left int...

Страница 38: ...Edit Tab have been described in the previous section 3 1 37 Monitoring rejection Continue to monitor the rejection situation by observing the small oval light at the bottom of the screen Green indica...

Страница 39: ...ter the required markers have been placed The SN10 Wave V trough marker must be placed for calculation of the waveform s signal to noise ratio displayed in the Recorded Curve Conditions table 3 1 39 D...

Страница 40: ...taken from the database to provide the selection of the corresponding normative latency data 3 1 42 Deleting individual curves Right click on the curve s handle to Delete the curve This will permanen...

Страница 41: ...hen entering the software in new test mode It is not available if you have entered the edit mode directly from the database 3 1 46 Merging curves creating a summed curve Curves of the same intensity s...

Страница 42: ...cond curve and select Add this curve The added curve will be indicated by an a following the intensity on the handle e g 60a Original curves 60a is the sum of the two other curves The added curve can...

Страница 43: ...The following information is displayed 1 Measured the number of sweeps recorded 2 Rejected the number of sweeps rejected 3 Rejection the rejection level used during testing 4 Masking indicates the mas...

Страница 44: ...right hand corner by clicking inside the box and then entering text These comments will be displayed only when the corresponding curve is selected All curve comments will be printed on the Curves Cond...

Страница 45: ...ove the double cursor by left clicking with the mouse 3 1 54 Signal to noise ratio calculation 3 1 Calculate the signal to noise ratio SNR of the waveform by placing the wave V marker and the SN10 wav...

Страница 46: ...nts of the Latency Tab 3 2 1 Latency values Select Latency to display the latencies of all marked waveforms for left and right ears 3 2 2 Interlatency values Select Interlatency to display the interpe...

Страница 47: ...data has been added in the System Setup Gender and age data are taken from the database to provide the selection of the corresponding normative latency data 3 2 4 Disabling desktop composition EPx5 so...

Страница 48: ...h follow these instructions 1 From the General setup tab change the Level measure method to Peak to trough 2 Press OK to save The Tr buttons will now appear in the Edit tab for manual positioning of t...

Страница 49: ...0 Enable Talk Forward Ctrl Shift F4 Show all Contra Curves Ctrl L Toggle to Latency Tab Ctrl R Toggle to Record Tab Ctrl E Toggle to Edit Tab Ctrl P Print session Shift F1 Context help Ctrl N Save New...

Страница 50: ...Risk of strangulation Keep cables away from baby s neck Avoid contact between the unused electrodes and any other conductive parts 3 5 1 Impedance check Press the Imp button and turn the dial fully cl...

Страница 51: ...pediatric or difficult to test populations A typical montage for an ABR threshold recording is RED Right mastoid or ear lobe reference or inverting BLUE Left mastoid or earlobe reference or inverting...

Страница 52: ...ction factors is available in some fitting software such as DSL v5 0a and Oticon s fitting software These offer a solution for the audiologist when the hearing thresholds supplied to them are not corr...

Страница 53: ...e must be observed 3 7 2 Editing of neuro latency recordings A Neuro latency recording is typically performed using Click and CE Chirp LS using stimuli level from 80 dB nHL Changes in response latency...

Страница 54: ...rrecting the average degree of recruitment 4kHz loss 50dB nHL no correction Deduct 0 1 ms per 10dB above 50dB nHL Selters Brackmann 1977 also described in NHSP UK Please refer to the Neuro Latency fun...

Страница 55: ...High forehead or CZ active or non inverting electrode BLACK Ground at low forehead or ipsilateral on implant side to reduce coil interference 3 8 2 Editing an eABR recording The electrical stimuli ar...

Страница 56: ...R test each electrode band are not tested due to the lengthy procedure of testing the many electrode bands Instead interpolation is used across the recorded electrode bands Below is an example of the...

Страница 57: ...trodes be used to measure the electrocochleogram While transtympanic electrodes will result in the most robust response it may not be feasible in many clinics 3 9 1 ECochG electrode montages TipTrode...

Страница 58: ...for Use EN Page 50 3 9 2 Editing an ECochG recording Example of Marked Points for Amplitude Ratio Example of Marked Points for Area Ratio Further information about ECochG recordings can be obtained fr...

Страница 59: ...ation as possible so electrodes are often placed in the ear canal at the eardrum TipTrode or TMtrode or by use of a trans tympanic electrode Electrode Montage of TipTrode and TMtrode is described unde...

Страница 60: ...imuli 3 11 1 Example of electrode montage AMLR RED Right mastoid or ear lobe reference or inverting BLUE Left mastoid or earlobe reference or inverting WHITE True CZ or vertex high forehead active or...

Страница 61: ...LR separate them from the cochlear implant stimuli artifacts seen under the traditionally eABR AMLR may this way be used to assess the efficacy of cochlear implant in activating the auditory pathway P...

Страница 62: ...ntion to stimuli the N1 increases up to 50 while P2 appears to diminish with increased attention by the subject on the signals The response habituates so it is important to limit the test session and...

Страница 63: ...est level response 5uV is threshold 3 12 4 Electrophysiological threshold estimation and infant hearing instrument fitting ALR threshold behavioral correction factors to estimated hearing thresholds t...

Страница 64: ...HITE True CZ active or non inverting BLACK Ground at cheek or low forehead distance of few cm to WHITE electrode must be observed The arousal state Patient attention of the patient has an effect on th...

Страница 65: ...e jumped electrodes Tip of nose of averaged reference jumped electrodes Filtering 1 30Hz 1 30Hz Analysis time Pre stimuli 100ms Post stimuli 700ms or more Pre stimuli 50ms or more Post stimuli 400ms o...

Страница 66: ...tem performance loop back LBK15 testing The Loop Back LBK15 box allows for an easy check of the impedance measuring system the stimulus quality and data acquisition Further information about LBK 15 pr...

Страница 67: ...start testing and provide accurate results 4 Only electrode gel intended for electroencephalography must be used Please follow the manufacturer s instructions regarding the use of the gel 5 The prese...

Страница 68: ...reen the stimulus is presented to the patient s ear and responses are recorded When the EMG contraction falls above or below the defined range the bar will appear in red for right ear and blue for lef...

Страница 69: ...and select Remove as augmented VEMP partner The VEMP asymmetry is calculated using the following formula Where LA is the larger of the left or right curve s amplitude SA is the smaller of the left or...

Страница 70: ...ontage for cVEMP Red Right SCM reference or inverting White Left SCM Black Ground Blue Clavicular joint active or non inverting 4 6 2 Stimuli for cVEMP Typically a 500Hz tone burst is used for cVEMP t...

Страница 71: ...oximately 150 sweeps per waveform shall be collected EMG controlled stimulus cVEMP only 50 V RMS to 150 V RMS subject to aim for 100 V 4 6 4 Procedure 1 Confirm low impedances below 5kOhm 2 Choose the...

Страница 72: ...for Use EN Page 64 4 6 6 Example of cVEMP result Example of a cVEMP recording in a patient with a lowered VEMP threshold Further information about the cVEMP procedure and recordings can be obtained fr...

Страница 73: ...ts 250 Hz 4 kHz Click Custom WAV file when Research Module license is enabled 4 7 3 Default collection parameters The standard oVEMP protocol settings for the stimulus and recordings are Stimulus para...

Страница 74: ...ir head 3 Start the test 4 7 5 Edit oVEMP results Use the Edit sheet to mark the peaks To select the VEMP partner double click the waveform handle then right click the waveform handle of the opposite...

Страница 75: ...arch files and folders go to help and open the folder of interest 5 1 1 Logging of each sweep for a later replay First set a folder to save the logged data in the protocol from where data needs to be...

Страница 76: ...t in the General Setup where to save the exported waveforms e g C Export waveform To export a specific curve of interest right click at the intensity handle and select to export this averaged waveform...

Страница 77: ...are made on an Eclipse with license for the research module function 5 1 5 Importing WAV files for stimuli More detailed information about calibration of WAV files can be found in the Eclipse Service...

Страница 78: ...0dB and use the slider to adjust the calibration 3 If needed perform the peSPL nHL correction select 0dB and use the slider to adjust the correction a Stimuli can also be saved and presented to use on...

Страница 79: ...ed stimuli open the software and the ABCDEF stimuli appears on the dropdown list 6 You can create your protocol using the new stimuli and run e g an LBK test to see the stimuli being recorded Please s...

Страница 80: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 72...

Страница 81: ...canal 5 Keep the box of ear tips outside the reach of the patient 6 Be sure to insert the OAE probe tip in a way which will assure an airtight fit without causing any harm to the patient Use of a corr...

Страница 82: ...e housing with water or insert non specified instruments into the OAE probe 6 1 Handling and selection of ear tips When using the Eclipse probe SanibelTM OAE ear tips must be used The SanibelTM ear ti...

Страница 83: ...ert the probe tip without an ear tip attached into the 2cc cavity provided with the Eclipse 3 Press the Start button and let the test run until it stops after 2 minutes Do not stop the test manually 4...

Страница 84: ...nd redo the probe test If the probe functions correctly after removing wax or debris continued use of the probe to test patients is ok If the probe status indicates Too noisy this indicates that the e...

Страница 85: ...tabase please see the instruction for use for OtoAccess 7 2 2 Starting from Noah 4 Ensure that the Eclipse is connected before opening the software module If the hardware is not detected the IA OAE Su...

Страница 86: ...ics about the error message and extra information can be added by the user outlining what they were doing before the crash occurred to assist in fixing the problem A screen shot of the software can al...

Страница 87: ...ffect after the IA OAE software is reopened Menu Setup Simulation mode activate the simulation mode and trial out the protocols or see how data is visualized Menu Print preview print print wizard and...

Страница 88: ...ed by ENT or medical specialist Probe performance is crucial to OAE test results We recommend that you conduct a probe test at the beginning of each day before starting to test on patients to ensure t...

Страница 89: ...indicates the DP points there have met the DP criteria as set in the protocol as show above for the right ear The DP points having no checkmark did not meet the DP criterion and can be either below t...

Страница 90: ...r to the Additional Information document for more details about the print wizard Print to PDF icon appears when setup via the General Setup This allows for printing directly to a PDF document that is...

Страница 91: ...is detected the status bar is grey Summary view toggles between displaying result graph or result graph with test summary tables Monaural Binaural view toggles between showing the results of a single...

Страница 92: ...st is stored in the memory Pause becomes active after testing has started It allows for pausing during a test START and STOP button are used for starting and stopping the session The probe check graph...

Страница 93: ...nt has ended without reaching the specified criterion for the individual point within the time allowed In the general setup it can be chosen whether or not this type of indication is shown The noise f...

Страница 94: ...oint to re test by right clicking on it A black circle around the measurement point indicates that it is selected After pressing Continue where the Start button was the selected point will be tested w...

Страница 95: ...ity percentage The Detected column displays a check mark when the specified frequency has met the criteria as setup in the protocol DP level Noise and SNR values are rounded based on raw data Therefor...

Страница 96: ...ted by ENT or medical specialist Probe performance is crucial to OAE test results We recommend that you conduct a probe test at the beginning of each day before starting to test on patients to ensure...

Страница 97: ...e top of the graph The checkmarks indicates the TE bars there have met the criteria as set in the protocol as show above for the right ear The TE bars having no checkmark did not meet the TE criterion...

Страница 98: ...r to the Additional Information document for more details about the print wizard Print to PDF icon appears when setup via the General Setup This allows for printing directly to a PDF document that is...

Страница 99: ...summary tables Monaural Binaural view toggles between showing the results of a single ear or both ears Report editor button opens a separate window for adding and saving notes to the current or histo...

Страница 100: ...lows for pausing during a test START and STOP button are used for starting and stopping the session The SNR Signal to noise ratio is displayed within each of the frequency bands tested and is calculat...

Страница 101: ...Only the waveform within this range accounts for the calculation of the waveform reproducibility percentage Clicking on the aqua or purple circles at the end of each waveform and moving the mouse all...

Страница 102: ...outside the tolerance set in the protocol setup the stimulus levels outside tolerance pop up dialogue will appear on screen Press Stop to stop testing Press Skip if you want to continue testing withou...

Страница 103: ...1 Categories you can select Templates to show all available templates Factory defaults to show only standard templates User defined to show only custom templates Hidden to show hidden templates My fa...

Страница 104: ...template and select View Show 10 Select one of the existing templates and press the My Favorites button to mark the template as a favorite The template can now be quickly found when My Favorites is se...

Страница 105: ...the system is ready to start testing and provide accurate results 4 Only electrode gel intended for electroencephalography must be used Please follow the manufacturer s instructions regarding the use...

Страница 106: ...from red to green as the impedance is found for each electrode Press the Imp button to exit the impedance mode before testing 8 4 Loop back LBK15 testing The LBK15 unit is not valid for performing a...

Страница 107: ...the print report 4 Select Print setup to select printer settings 5 Select Exit to exit the EP15 25 software Edit provides access to delete right delete left and delete right left 1 Select delete righ...

Страница 108: ...er of the screen 8 5 4 Printing The Print icon prints the report for the selected session 8 5 5 Electronic help Click on the Electronic Help icon and then point click on an item you require more infor...

Страница 109: ...88 B 2019 03 Eclipse Instructions for Use EN Page 101 8 5 9 Test status window Displays the Pass criteria status and the Recording time during and after testing 8 5 10 Ear Indicates which ear is being...

Страница 110: ...he rejection level by using the arrows above the EEG graph 8 6 3 EEG noise bar The EEG Noise Bar can be displayed instead of raw EEG Graph When the noise is below the black line the VU meter is green...

Страница 111: ...a pass indication is displayed in the status bar above the curve If the 100 detection line is not reached during the 120 seconds a Refer is displayed 8 7 PC shortcuts Shortcut Description F1 Help F2...

Страница 112: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 104...

Страница 113: ...cracks and electrodes e g check the expiry date of the disposable electrodes check cables for damage to verify that the system is ready to start testing and provide accurate results 4 Only electrode...

Страница 114: ...ground connection black can be placed on the low forehead or side of the forehead The placement of the ground electrode is not very critical Note that all four electrodes must be positioned The electr...

Страница 115: ...place is much easier to rub down with the skin abrasive gel easier than the cheek which is softer Please note that even though the impedance checking system is designed to give a direct indication of...

Страница 116: ...is evaluated to determine rejection 9 5 Impedance check Press the Imp button and turn the dial fully clockwise Turn the dial slowly counter clockwise The LED for each electrode will change from red t...

Страница 117: ...es that are available for print are displayed 4 Select Print setup to select printer settings 5 Select Exit to exit the EP15 25 software Help provides access Help topics and About 1 Select Operation M...

Страница 118: ...ll not be saved To exit without saving click on the red X in the upper right hand corner of the screen This icon is not available through Noah 4 When clicking on Exit from the File menu or exiting via...

Страница 119: ...tructions for Use EN Page 111 9 7 10 Stimulus rate Select the appropriate stimulus rate 40Hz or 90Hz for testing To change during testing when the patient state changes all test frequencies must have...

Страница 120: ...rows between the two raw EEG curve graphs During testing the rejection level can only be adjusted within certain limits To change the rejection level outside these limits the test must be stopped 9 8...

Страница 121: ...the curve and provide numerical information for the probability percentage and time at any point of the curve 9 8 5 ASSR table The ASSR table displays the detection curve information in a table format...

Страница 122: ...Adjusting the stimulus intensity Right click on the Right and Left text above the Test Frequency Graphs to select and initiate an overall stimulation intensity for the entire selected ear immediately...

Страница 123: ...0120588 B 2019 03 Eclipse Instructions for Use EN Page 115 9 8 8 Stopping a test frequency intensity Right click within the desired Test Frequency Graph to Stop the Current Ongoing Intensity Immediate...

Страница 124: ...e following section describes the elements of the Audiogram Tab The Audiogram Tab is where the resulting estimated audiogram is displayed and correction factors can be modified 9 9 1 Estimated audiogr...

Страница 125: ...agging it to the desired position with the mouse The initial estimate based on the selected correction factor table the grey audiometric symbol is displayed and saved in the database 9 9 2 No response...

Страница 126: ...ion Manual F2 Start Stop test F4 Pause Resume test F7 Report F8 Print session Ctrl F7 Temporary Setup Ctrl P Print session Alt X Save Exit Page down Toggle backwards through historic sessions Page up...

Страница 127: ...onnectors and that it is not exposed to any kind of mechanical load which could involve damage 6 For maximum electrical safety turn off the power from the mains when the instrument it is left unused 7...

Страница 128: ...the instrument by wiping the outer case with a lint free cloth lightly dampened in cleaning solution Clean cushions and the patient tone switch and other parts with a lint free cloth lightly dampened...

Страница 129: ...h and thin rigid plastic cord 5 Use the plastic cord or brush to remove any debris from the probe tip Always insert the cleaning tool from the rear to avoid accumulation of debris inside the small rou...

Страница 130: ...ser Accessories are free from defects in material and workmanship under normal use and service for a period of ninety 90 days from the date of delivery by Interacoustics to the first purchaser If any...

Страница 131: ...n to assume on behalf of Interacoustics any other liability in connection with the sale of Interacoustics products INTERACOUSTICS DISCLAIMS ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING ANY WARR...

Страница 132: ...D 0120588 B 2019 03 Eclipse Instructions for Use EN Page 124...

Страница 133: ...wer Supply Input Volts 100 240VAC 50 60Hz Consumption 26W 0 3A Max Safety marking ECM60UT31 XE0410 and cURus Operating environment Operating Temperature 15 35 C 59 95 F Rel Humidity 30 90 Ambient Pres...

Страница 134: ...each electrode Measurement frequency 33 Hz Waveform Rectangular Measurement current 19 A Range 0 5 k 25 k Stimulus Stimulus rate 0 1 to 80 1 stimuli per second in steps of 0 1 Envelopes Windows Bartl...

Страница 135: ...er without wave peak latency displacement High Pass Filter 0 83 Hz to 500 Hz depending on the measurement type DSP Low Pass Filter 100 300 750 1k 1 5k 2k 3k 4k 5k 7 5k Hz DSP High Pass Filter 0 5 1 0...

Страница 136: ...25 47 5 2000 11 5 9 31 3000 30 25 5 46 3000 13 11 5 30 4000 32 5 27 5 52 4000 15 12 35 5 ISO 389 6 2007 ISO 389 1 2000 ISO 389 2 1994 ISO 389 3 1994 Click ECochG ABR15 ABR30 AMLR Neuro VEMP 0 dB Clic...

Страница 137: ...g Analysis time 5 seconds to 30 minutes Sampling frequency 30 kHz A D Resolution 16 bit 3 7 Hz resolution Artifact Reject System 0 to 60 dB SPL or off Applicable during testing SNR Criteria Adjustable...

Страница 138: ...L or off Applicable during testing Stimulus Tolerance Adjustable between 1 and 10 dB SNR Criteria Adjustable between 3 and 25 dB Probe check window 256 points frequency response of the ear canal due t...

Страница 139: ...18 dB Typical 130 dB 100 Hz Radio frequency immunity Typically 25 dB improvement over previous available designs Max input offset voltage 2 5 V Input impedance 10 M 170 pF Power from main unit Insulat...

Страница 140: ...endent calibration optional Bone conductor optional Level 0 100 dB nHL in 5 dB steps NB CE Chirp Freq 500 1000 2000 and 4000 Hz both ears same time Bandwidth 1 octave octave 3 dB Masking White noise 0...

Страница 141: ...result in improper operation The list of accessories transducers and cables can be found in the EMC appendix of this instruction Portable RF communications equipment including peripherals such as ant...

Страница 142: ...Rated Maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 17 P 80 MHz to 800 MHz d 1 17 P 800 MHz to 2 5 GHz d 2 23 P 0 01 0 12 0 1...

Страница 143: ...s W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a shou...

Страница 144: ...48A Bone Conductor Radio Ear Bone OAE Probe Interacoustics Opt25 Cochlear Nucleus Trigger Cable Interacoustics Cochlear Nucleus Conformance to the EMC requirements as specified in IEC 60601 1 2 is ens...

Страница 145: ...nal Narrow Band CE Chirp LS stimuli 0 5 1 2 4 kHz Optional x Optional Tone Burst stimuli 0 25 8kHz x x x Recording window 15 and 30 ms Up to 980 ms 150 ms ABR x x Rate Study x x x ECochG Optional x Op...

Страница 146: ...per frequency ABR3A DD45 TDH39 B71 B81 Stimulus short 2 1 2 Long short 2 1 2 Long short 2 1 2 Long short 2 1 2 Long short 2 1 2 Long Burst 250 105 115 105 110 105 110 50 50 50 55 Burst 500 110 120 115...

Страница 147: ...module Test types functionality DPOAE Module Stimulus level 30 70 dB SPL Stimulus range 500 10000Hz Test time Min 2 sec unlimited DP Gram x DP Input Output x Normative data display option x Checkmark...

Страница 148: ...d CE Chirp stimuli 0 5 1 2 4 kHz x Recording time Up to 15 min per curve Stimulus rate 40 or 90 Hz Transducer options Headphone DT48A Inserts Bone nHL to eHL correction factors Child Adult x Residual...

Страница 149: ...Returned according to agreement with Interacoustics Other Date Person Please provide e mail address or fax No to whom Interacoustics may confirm reception of the returned goods The above mentioned ite...