52
26 Error Codes
�
ERR01:
the weight does not reach stability after a tare operation
~
Protect the
balance from air flows or from vibrations of the working table.
�
ERR02:
impossible to start the calibration operation due to balance instability
~
Protect the balance from air flows or from vibrations of the working table.
�
ERR03:
calibration weight not correct or balance unstable
~
Calibrate with correct
weight or protect the balance from environment disturbs.
�
ERR04:
weight of samples for the piece counting function not adequate or unstable
~
Select a bigger number of samples or protect the balance from vibrations.
�
ERR05:
impossible to print due to instability
~
Protect the balance from
environment disturbs.
�
ERR07:
error in inserting the data.
�
“UNLOAD”:
weight loaded on the pan or pan not positioned properly
~
Remove
the weight from the pan or position properly the pan and underpan.
�
“CAL But”:
the balance requires to be re-calibrated
~
Unload weights, if any, on the
pan, and press the CAL button.
�
: Overange condition
~
Unload the weights loaded on the
pan.
�
: Underange condition
~
Place properly pan and underpan.
27 Maintenance and care
Regular maintenance of yours balance guarantee accurate measurements.
�
Cleaning
Before cleaning the balance unplug the power supply of the balance from the
voltage supply of your room. Do not use aggressive cleaning product (as solvents or
similar), use a humid towel with soft detergent, Avoid liquids to go inside the
instruments during the cleaning. Wipe the balance with a soft towel. Parts of
samples or powder can be removed using a brush or vacuum cleaner.
�
Safety checks
Safety of the instrument is no more guaranteed when:
balance power supply is clearly damaged
balance power supply is not working anymore
balance power supply is stored for long time in hard environment conditions. In
these instances refer to the assistance centre where specialized technician will
make reparations to bring back the instrument in the safety conditions eventually.