4
1 Installation instruction
WARNING:
Please follow carefully these steps for installing and use the new balance
before starting your work routine. Failure to operate the unit in accordance
with these guidelines will affect the weighing results.
�
Remove
the balance and accessories from the cartoon and check for any visible
damage of the instrument.
�
Do not install
the balance in a place with air flows, heavy thermal changes and
vibrations.
�
Do not use
the balance in explosion risk environment.
�
The humidity rate
of the balance environment must be between 45% and 75%
�
Place
the weighing pan and the support pan on the balance (see pag. 5).
�
Level
the balance using the level bubble and levelling feet located underneath the
case
�
Connect
power supply to connector 2 located on the rear panel of the unit (see
pag. 7).
�
Insert
power unit into AC outlet,
which should be easily accessible;
after few
seconds the balance will automatically switch on.
�
Wait 30 minutes from switch on
and then calibrate the balance using the
appropriate mass, following the instructions (pag.9-10)