• Chapter 6 • Disassembly and Sterilization of CUSA
®
NXT After Surgery
8.
Turn off the power on the rear panel of the CUSA
®
NXT
Console and
Service Module (if utilized). Remove the main power cord from the
Console or Service Module as appropriate, wipe clean and store in the
rear cupboard of the Service Module (if utilized).
9.
Disconnect the Footswitch and wipe clean. If the Footswitch is
contaminated with blood or fluid, it may be cleaned by immersing it in
warm water containing detergent or disinfectant (the port must be kept
dry) and rinsed. Allow the Footswitch to drain after washing and store in
the rear storage area of the Service Module (if utilized).
10.
Wipe the Console and the Service Module with a clean cloth dampened
with warm water and a mild detergent, disinfectant or alcohol when
necessary to remove blood or dirt. Lower the irrigation pole by turning the
knurled locking ring counter-clockwise.
11.
Remove the handpiece cable from the Console and disconnect from the
handpiece. Wipe the cable with a clean cloth dampened with warm water
and a mild detergent, disinfectant or alcohol when necessary to remove
blood or dirt.
Notice
Do not remove the aspiration filter from the designated port prior to powering off
the system.
Notice
Do not re-use the tubing kit as this compromises sterility and may lead to
damage of the Console and/or Service Module.
Notice
Do not remove the contamination guard situated in the receptacle bay at the
front of the Service Module.
Warning
Electric Shock Hazard – Always unplug the CUSA NXT System before cleaning
it.
Warning
Do not allow fluids to enter the Console or Service Module.
Notice
When dismantling and assembling handpieces of different operating
frequencies (24 kHz or 35 kHz), care should be taken to ensure that the correct
tips are fitted to the correct handle section. This is achieved by following the
color coding, bronze for 24 kHz and gold for 35 kHz. Incorrectly fitted
handpieces will either be non-functional or may not perform as intended.
Содержание CUSA NXT
Страница 3: ...This page is intentionally left blank...
Страница 7: ...iv This page is intentionally left blank...
Страница 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Страница 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Страница 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Страница 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Страница 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Страница 113: ...106 Index This page is intentionally left blank...