Chapter 3 • Using the Console, Service Module, and Handpieces •
37
•
Fast Flush
(right, blue).
To connect the Footswitch to the Console, attach the Footswitch cord to the
appropriate port on the rear of the Console panel. Align the red dot on the
Footswitch plug with the red dot on the Footswitch connection point, and
gently insert the plug until you hear a clicking sound.
Note that the Footswitch is internally sealed and may be cleaned by
immersion (see page 74).
Pressing the Power Pedal
Pressing this pedal will:
•
Switch on the irrigation at the preset rate indicated on the front panel
irrigation display.
•
Switch on the ultrasonic power to a level which varies in an approximately
linear fashion from zero (when the pedal is not pressed), to the maximum
value indicated on the front panel power display (when the pedal is fully
pressed).
Notice
After releasing the
Power
pedal, the pinch valve operates momentarily to block
aspiration and facilitate tissue release from the handpiece tip.
Notice
Do not press the ultrasonics Power pedal on the Footswitch during system
startup or while connecting components to the Console.
Power
Fast Flush pedal
pedal
Содержание CUSA NXT
Страница 3: ...This page is intentionally left blank...
Страница 7: ...iv This page is intentionally left blank...
Страница 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Страница 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Страница 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Страница 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Страница 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Страница 113: ...106 Index This page is intentionally left blank...