Chapter 2 • CUSA
®
NXT System Setup and Components •
29
Tightening the Handpieces with the Wrench Kits
Clearing Blocked Handpieces
Inspecting Handpiece Tips for Damage
Priming the Handpieces
Caution
Only use the CUSA Selector wrench kit to tighten tips and extensions on
the transducer of the handpiece. Always secure the handle of the
handpiece in the holder provided. Never dismantle or assemble the
handpiece without placing the handle section of the handpiece in the holder.
Failure to comply with this instruction could result in permanent damage to
the transducer and will invalidate the warranty.
Caution
Do not attempt to clear a blocked handpiece by passing the stylet down the
rear inlet of the handpiece. This may result in damage to the handpiece.
Always pass the stylet from the tip end.
Warning
Do not use a damaged handpiece with the CUSA NXT System.This may result
in injury to the patient or surgical personnel. A damaged handpiece must be
returned to the manufacturer for repair. Do not attempt to dismantle the rear
handle sections of the handpieces; it does not contain any serviceable parts.
Caution
Do not use a
handpiece with a broken tip. Inspect the tip for damage prior to
use. If the tip is damaged it may be replaced by the operator.
Caution
Do not use dry air sterilization methods involving temperatures in excess of
139 °C (282 °F); this may damage the handpiece.
Caution
Do not power up the CUSA Selector handpiece without fully priming the
handpiece with irrigant (i.e. irrigant must flow from the tip of the handpiece).
Failure to prime the handpiece could result in permanent damage to the
handpiece (internal and/or external components) and may induce heat
exchange with the patient or user, and cause tissue burns.
Содержание CUSA NXT
Страница 3: ...This page is intentionally left blank...
Страница 7: ...iv This page is intentionally left blank...
Страница 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Страница 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Страница 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Страница 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Страница 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Страница 113: ...106 Index This page is intentionally left blank...