Chapter 4 • Preparing the CUSA
®
NXT for Surgery •
61
3.
Retain the yellow stylet and white hose clips in the sterile field.
4.
Connect the tubing to the handpieces. Attach the double lumen tube from
the standard tubing kit to the irrigation (
1
) and aspiration (
2
) ports on the
back of the handpiece.
If you are using the
MicroSurgical
handpieces:
• For the
MicroSurgical
handpiece, attach the micro tubing kit to the
irrigation (
1
) and aspiration (
2
) ports on the back of the handpiece.
Connect the other end of the tubing to the standard tubing kit.
5.
Using the white hose clips supplied with the tubing kit, secure the tubing
and handpiece cord.
6.
Depending on the Console location, reserve the desired amount of tubing
to be used in the sterile field.
7.
Hand the remaining tubing to the Console location in the non-sterile area.
Notice
All of the handpiece and tubing kit components in the CUSA NXT System are
not manufactured with dry natural rubber or natural rubber latex.
2
1
Back End of Handpiece
Содержание CUSA NXT
Страница 3: ...This page is intentionally left blank...
Страница 7: ...iv This page is intentionally left blank...
Страница 39: ...32 Chapter 2 CUSA NXT System Setup and Components This page is intentionally left blank...
Страница 73: ...66 Chapter 4 Preparing the CUSA NXT for Surgery This page is intentionally left blank...
Страница 79: ...72 Chapter 5 Operating the CUSA NXT During Surgery This page is intentionally left blank...
Страница 95: ...88 Chapter 7 Troubleshooting the CUSA NXT System This page is intentionally left blank...
Страница 105: ...98 Appendix A Technical Specifications This page is intentionally left blank...
Страница 113: ...106 Index This page is intentionally left blank...