background image

Gain calibration 

Enter test gas concentration (value between 30 and 70 % of the measuring range)

The current sensor element sensitivity is read with "Read".

Slide calibration adapter carefully onto the Sensor Cartridge.

Apply test gas (flow rate according to the table "calibration", 1 bar ± 10%) to the Sensor Cartridge.

When the value is stable, the new gain factor is calculated with "Calibration".

The new gain factor is checked for plausibility and stored in the buffer memory. The current measured value 
is output with the new gain factor and the sensor element sensibility is updated.  

With "Save" the new gain factor is written in the SC memory, only then the gain calibration has been
successfully completed. If you exit the menu without pressing "Save", the original gain data for the
measured value calculation will continue to be used.

By limiting the gain factor, calibration will not be possible anymore when the sensitivity of the sensor 

reaches a residual sensitivity of 40 %. Then the Sensor Cartridge has to be replaced  

For more information, see the user manual of the 

DPT6 EasyConfig Software

4.6 Exchange of Sensor Cartridge 

Instead of the on-site calibration, the used SC can be easily and conveniently replaced by a calibrated one. 

The communication of the local bus (Sensor Cartridge <> BSB) is continuously monitored during 
operation and results in  an immediate error message on the gas controller in case of fault or 
interruption. When replacing the sensor unit, the communication of the local bus is also 
interrupted when unplugging the SC connector which leads to an immediate triggering of the 

error message. 

Disconnect the SC connector from the BSB or the RSB (error message will be activated).

Loosen the locknut.

Remove used SC.

Take calibrated SC out of the  original packaging, check for gas type, measuring range and valid
calibration date.

Insert the SC and retighten with lock nut.

Insert the SC plug into the socket at the BSB or RSB. Check plug for proper engagement.

The local bus communication is automatically established and tested. At the same time the gas type and the 
measuring range of the "new" SC are compared with the data stored in the  BSB. If they match and the 
communication is correct, the error message will be automatically acknowledged at the Gas Controller.  

The yellow LED of the BSB flashes with a pulse duration of 1 sec., while the SC connector is disconnected 
(communication error). After the local bus communication has been re-established and the conformity test 

has been successful, the LED will go into flashing mode with 3 sec. pulse duration.  

Perform functional  test of the exchanged SC with gas generator. On successful completion of the
test, the flashing LED goes off after approximately 10 sec.

The sensor element is treated with a defined gas concentration with the help of the gas generator. As a result 
the measurement signal acknowledges the LED when an internal switching threshold is reached. With this 

test, the complete function chain “Sensor  Element > Sensor Cartridge> Local Bus> BSB>  Field Bus> GC 
Controller” is tested. 

INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com

Specifications subject to change without notice. | GASB2_03_E_0220 | USA 200306 | Page 8 of 15

DT6 – UserManual

Содержание PolyGard 2 DT6

Страница 1: ...Transmitters Serial No SB 03 User Manual Version February 2020 en ref optional remote sensor DR6 INTEC Controls 12700 Stowe Drive Suite 100 Poway CA 92064 Ph 858 578 7887 888 GO INTEC inteccontrols c...

Страница 2: ...5 Inspection and Service 9 5 1 Inspection 9 5 2 Service and Calibration 9 6 Troubleshooting 10 6 1 DT6 Basic Sensor Board 10 6 2 SC2 Sensor Cartridge Messages at the Tool Controller 10 7 Technical Dat...

Страница 3: ...local bus level is done via the STL6 Service Tool or with the DPT6 EasyConfig Software which is directly connected to the BSB The cable topology for the RS 485 field bus can be taken from the Guideli...

Страница 4: ...eight depends on the relative density of the gas type to be monitored Choose mounting location of the sensor according to the local regulations Consider ventilation conditions Do not mount the sensor...

Страница 5: ...SB connect cable according to connection diagram fig 1 Replug terminal carefully on the BSB at X4 For remote sensor Insert and connect local bus cable at the basic and remote sensor board Remove termi...

Страница 6: ...as type and defining the measurement range the SC is connected to the input Up to two different SC can be connected at the DT6 with a third to an attached remote DR6 The physical position of the SC do...

Страница 7: ...on you have to activate the mode Special Mode at the BSB only then the calibration menu is enabled During the special mode the BSB doesn t issue alerts Select the Sensor Cartridge to be calibrated by...

Страница 8: ...g the SC connector which leads to an immediate triggering of the error message Disconnect the SC connector from the BSB or the RSB error message will be activated Loosen the locknut Remove used SC Tak...

Страница 9: ...be documented The date for the next maintenance has to be affixed to the sensor 5 1 Inspection Gas sensors should be controlled regularly by a competent person according to EN 45544 4 The following ha...

Страница 10: ...h with registered ones Check SC data registration data for conformity Message at the Tool Controller 24 V DC voltage range 5 V DC voltage range Temp range WatchDog triggered Internal error Replace BSB...

Страница 11: ...B x D 94 x 130 x 57 mm 3 7 x 5 1 x 2 2 in Weight 0 3 kg 0 7 lb Protection class delivery status NEMA 4X IP 65 Installation Wall mounting Wire connection Field bus Local bus for SC Screw type terminal...

Страница 12: ...oard Local Bus only for Remote Sensor Board Field Bus 1 2 3 4 0 VDC 24 VDC Bus_A Bus_B Bus_B Bus_A 24 VDC 0 VDC Analog Output 4 20 mA 0 VDC Field Bus 1 2 3 500 Remove resistor if device is connected X...

Страница 13: ...rd with SC X 2 X 2 X 2 X 2 X 2 X2 X 3 X 3 X 3 SB_03 SB_03 SB_03 SB_03 SB_03 SB_03 X4 4 3 2 1 X4 4 3 2 1 X4 4 3 2 1 X7 3 2 1 X7 3 2 1 X7 3 2 1 X7 3 2 1 X7 3 2 1 X7 3 2 1 Basic Sensor Board Basic Sensor...

Страница 14: ...Mounting Mounting Mounting Mounting 115 mm 130 mm 180 mm D 4 mm INTEC Controls 12700 Stowe Drive Suite 100 Poway CA 92064 Ph 858 578 7887 888 GO INTEC inteccontrols com Specifications subject to chan...

Страница 15: ...de ANSI NFPA70 The equipotential bonding required also e g secondary potential to earth or grounding measures must be carried out in accordance with the respective project requirements It is important...

Отзывы: