5.2 Manual Zero Point
•
The zero-point is factory-set; therefore zero calibration is not necessary.
5.3 Gain
•
Connect calibration adapter carefully to the sensor element.
•
Apply calibration test gas O
2
(150 ml/min; 1 Bar (14.5 psi) ± 10%).
•
Wait two minutes until the signal is stable, adjust control voltage with potentiometer ”Gain” until the signal
corresponds to the calculated value ± 2 mV, see “Calculation of Control Voltage”.
•
Remove calibration adapter with a careful light turn. Check the sensor for correct mounting!
By limiting the gain factor, calibration will not be possible any more when the sensitivity of the sensor
reaches a residual sensitivity of 30 %. Then the sensor has to be replaced.
Calculation of Control Voltage
Signal start 2 V / 4 mA
Control voltage (mV) = 160 (mV) x test gas concentration O
2
(vol. %) + 40 (mV)
measuring range O
2
(vol. %)
Signal start 0 V / 0 mA
Control voltage (mV) = 200 (mV) x test gas concentration O
2
(vol. %)
measuring range O
2
(vol. %)
Example:
Measuring range
25 vol. %
Test gas concentration
20
.8
vol. % O
2
Control voltage: Signal start 2 V / 4 mA
173.
1
mV
Control voltage: Signal start 0 V / 0 mA
16
6
.
4
mV
Signal start 0 V / 0 mA
Signal start 2 V / 4 mA
160 (mV) x 20.
8
(vol. %) + 40 (mV) = 173.
1
mV
200 (mV) x 20.
8
(vol. %) = 16
6
.
4
mV
25 (vol. %)
25 (vol. %)
Specifications subject to change without notice. | USA 150109 | Page 6 of 13
12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.IN
TEC
| relevantsolutions.com/inteccontrols
AT-1195_V3 – UserManual
Polygard® is a registered trademark of MSR | AT-1195_V3_UserManual
Specifications subject to change without notice. | USA 160104 | Page 6 of 13
12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.IN
TEC
| relevantsolutions.com/inteccontrols
AT-1195 V3 – UserManual
PolyGard
®
is a registered trademark of MSR-Electronic GmbH | AT-1195_V3_UserManual
INTEC Controls | 12700 Stowe Drive, Suite 100, Poway, CA 92064 | Ph: (858) 578.7887 & (888) GO.INTEC | inteccontrols.com
Specifications subject to change without notice. | USA 200211 | Page 6 of 13
AT-1195 V3 – UserManual