48
3.3 INSTALLATION
Distance Order
Power up the remaining loggers, starting with those closest to the receiver – this order will minimise the
time necessary for them all to connect.
In most cases these loggers should establish a preliminary place in the mesh and connect within a few
minutes, otherwise the full meshing process will allow them to connect after 30 minutes.
Be Patient
It can sometimes take a while for a network to establish itself.
If a logger is not listing in Tinytag Explorer it may be tempting to reboot it or move it to another
location. Rather than helping, doing this may actually slow the connection of the logger; it is better to
be patient and give the system some time.
If a logger hasn’t signed in after a couple of hours, that would be the time to try rebooting the unit or
moving it to another location.
Transplanting a Network to a New Location
If moving a complete network to a new location, the following steps should be carried out:
• Set the loggers to a 2 minute logging interval to speed up the connection process (this can be
adjusted to a longer logging interval once the system has configured itself in its new location).
• Assuming the existing data is no longer required; clear the loggers of historical data before turning
them off (this means the system won’t be trying to download old data whilst trying to configure
itself).
• Power down all devices (including receivers) before starting over in the new location, as detailed
previously.
Содержание Tinytag Connect
Страница 1: ...Tinytag connect User Manual Tinytag Explorer Connect Software Included...
Страница 2: ......
Страница 5: ...Getting Started...
Страница 22: ...20...
Страница 23: ...System Information...
Страница 45: ...Radio...
Страница 63: ...Ethernet LAN...
Страница 70: ......
Страница 71: ......