Transport floor model frames
Product Support:
Warnings
Hazard - do not lift the frame by the base.
The base covers cannot support the weight of the frame. If you try to lift the frame with a
forklift under the base, the base covers could break and cause personal injury and
equipment damage.
Hazard - do not lift the frame by the top plate.
The top plate cannot support the weight of the frame. If you try to lift the frame with a
forklift under the top plate, the top plate could break and cause personal injury and
equipment damage.
Hazard - always use a lifting device or straps with sufficient capacity to lift the load
frame.
You must use a lifting device or straps rated for a capacity at least twice the weight of
the load frame.
Hazard - Do not tilt an unsupported load frame more than 10° when it is in the upright
position.
Tilting by more than 10° may topple the load frame and cause personal injury and
equipment damage.
Figure 21. Maximum tilt angle
10°
10°