background image

2

 2.  Function

100

0.025

L

HD

=

-

 The formula is as below:

 relative HD means it is types D Shore Hardness.

 According to this formula as above, the types D Shore hardness is 

 to the displacement of the pressed needle. Through measuring the 
 displacement of the pressed needle, the types D Shore Hardness 
 can be calculated. 

 1) Peak value latching, Average value calculation, Low-voltage alarming.

 2) Auto power off.

 In three minutes, if there isn't any operation, the durometer will be 
 auto power off. Automatically shut down before the buzzer rang five 
 times, to indicate that the machine will automatically shut down.

 3) Large LCD

 The high resolution industry class LCD is applied to this durometer. 

These make it convenience to read.

 4) Menu operation and management interface

The operation and management interface is menu type. It is easy to 
learn and operate.

 5) Fully data management

This durometer can storage, delete and print the test data. It also can 
set an interval to sorting samples and accomplish the basic statistics 
and calculations of the test data.

 6) The content to be printed can be customized.

 7) The large capacity rechargeable Ni-MH is built-in.

 3.  Application field

 This durometer is mainly used to measure the hardness of hard rubber 
 and hard plastic, such as: thermoplastic, hard resin, flooring, bowing. 
 Especially for on-site hardness measurement of rubber and plastic  
 products. 

15

2.  Print report

There are several print modes built-in the durometer. You can select by 
yourself.

3.  Auto-print

There are several print modes built-in the durometer. You can select by 
yourself.

1) Auto-print

 The auto-print command will be carried out during measuring if this 

 function is set to on. The durometer will send the data to the thermal 
 print after the finishing of one statistic group.

2) Concentration print

 The data can't be printed during measuring. You can print the data 

 

There are two options under this mode:  print selected data” 

later. 

 

all data” explained as above.

and “print 

Data storage 

This durometer can storage 500 group data at most (one group data 
include: one hardness value, the date/time/sample code). if 500 group 
data have been saved; no more data can be saved. You need to delete 
part or all data. If 500 group data have been saved, the screen will 
display “M FULL” under the measuring interface.

Hardness testing

Please press the needle on the sample and make sure the presser foot 
is contact closely to the sample. When the data is stable, the buzzer will 
make a sound and the value will show until the next measuring.
Current measuring mode: maximum average 
value mode
Current measured value: 58.1HD
Current statistic times: 5
The measured number: 2  
Current average value: 57.8HD
Peak difference: 0.7HD
Multi-point measuring will be realized by 
repeating above steps.

Fig 9-1

Содержание ISH-DSD

Страница 1: ...ISH DSD www insize com OPERATION MANUAL DIGITAL SHORE DUROMETER MN ISH DSD E V0...

Страница 2: ......

Страница 3: ...foot surface The value of L indicates the value of Shore Hardness of the sample Larger the value of L smaller the Shore Hardness Conversely smaller the value of the L sample s Larger the sample s Shor...

Страница 4: ...n flooring bowing Especially for on site hardness measurement of rubber and plastic products 15 2 Print report There are several print modes built in the durometer You can select by yourself 3 Auto pr...

Страница 5: ...Low voltage alarming and recharge The durometer will be recharged by the packaged charger The durometer will not be auto power on during recharging The charging status will show on the measuring inter...

Страница 6: ...e current times Measure mode Indicating the current measure mode AVE Average measure mode Max Maximum value measure mode MAX AVE Maximum average value measure mode 2 Menu 7 6 5 4 3 2 1 Fig 4 1 Hardnes...

Страница 7: ...r this menu All data will be deleted by selection of YES and this operation will be aborted by selection of NO Fig 4 22 5 Print setting The data saved in the durometer can be printed by the thermal pr...

Страница 8: ...ssing button UP The measure date time sample code of the current data can be shown pressing button OK 6 1 Calibration Pressing button OK to enter the menu select Calibration press button The durometer...

Страница 9: ...tically The average value of maximum value of every measuring will be calculated and this value will be taken as the last result Measure times setting The statistic times can be set under this menu Th...

Страница 10: ...l be button RETURN be aborted by press 8 Tolerance limit setting Under this menu the tolerance limit can be set When the measuring data is exceed the setting value the icon of will be showed on the LC...

Отзывы: