SCAMAX
®
2600/4000
Tech. Manual 06/2002
T-52
The initial feed belts and the separator rollers cover wear over a period of time. This results in the
distance between feed belts and separator rollers becoming ever greater. The setting on the BCD-switch
has to be decreased to compensate for the wear and tear. When a setting of '0' has been reached the
switch on the circuit board has to be changed from HIGH (factory setting) to LOW and a machine reset
has to be executed (switch scanner off, then on). The previous setting of '0' corresponds now to a setting
of '5'.
The HIGH/LOW switch also activates the following service routines; some of them are described in the
User Manual. Selection of the service routine is via the BCD-switch and initiation of the routine is via the
push buttons (see User Manual, for example white calibration):
BCD-value 9:
White Calibration
(see User Manual).
BCD-value 8:
Calibrating the Optical Double Feed Detector
(SCAMAX 2500 only).
BCD-value 7:
Cleaning Routine
(see User Manual).
BCD-value 6:
Test Scan Routine
. Facilitates testing of the mechanical functions of the
Scanner without the need for a PC to be connected to it.
The PC's control functions are simulated in the controller and any errors
Are indicated by the red LED in the start switch being illuminated.
Содержание SCAMAX 2600
Страница 7: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 7...
Страница 12: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 12...
Страница 14: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 14...
Страница 16: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 16 Illustration of optical reduction...
Страница 34: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 34...
Страница 49: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 49...
Страница 56: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 56...
Страница 62: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 62...
Страница 66: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 66...
Страница 67: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 67...