SCAMAX
®
2600/4000
Tech. Manual 06/2002
T-18
3.1.2 Camera Faults
The service program is used to determine whether a camera is faulty. The program can display the
camera status. If the camera status is shown as OK it is possible that the camera sends wrong data or no
data at all. This fact would be shown by an unexpected curve formation, heavy noise or absence of a
channel. If a problem is diagnosed it will not be immediately obvious whether it is caused by the camera
board and/or the CCD sensor. Therefore, several solutions present themselves. In the majority of cases
an exchange of the camera board will yield the result desired. This methodology has the advantage of
retaining the optical alignment of the camera.
3.1.2.1 Changing Camera Board Type 2+3
Switch off Scanner!
All activities have to undertaken in accordance with current ESD regulations!
Remove the two screws of the EMV camera housing and lift lid.
Pull the post connector from the board (1x for SCAMAX
®
2600
and 2x for SCAMAX
®
4000
) and un-screw
the four plastic nuts.
Using a pair of pliers pull the board backwards out of the housing by the earth straps. Ensure the board is
pulled squarely out of the housing; otherwise the CCD sensor's contact pins will be bent.
Carefully insert the new board and fasten using the plastic nuts.
Front
Back
Camera-EMV-Housing
Earth straps
Plastic nuts
Содержание SCAMAX 2600
Страница 7: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 7...
Страница 12: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 12...
Страница 14: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 14...
Страница 16: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 16 Illustration of optical reduction...
Страница 34: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 34...
Страница 49: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 49...
Страница 56: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 56...
Страница 62: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 62...
Страница 66: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 66...
Страница 67: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 67...