SCAMAX
®
2600/4000
Tech. Manual 06/2002
T-29
3.3 Exchanging the Controller Board
Switch off the Scanner!
All activities are to be conducted according to current ESD regulations!
Prior to exchanging the controller board the scan counter has to be read using the service program (menu
item: STATUS). As well, the parameter file (menu item PARAMETERS) has to be saved. The value of the
scan counter has to be written down. If communication is impossible due to a faulty controller, these
parameters can be obtained from InoTec by quoting the scanner’s serial number.
Action necessary for the changeover:
1.) Remove the left hand cover and the cover plate of the controller
2.) Remove all cables and the EPROM (FIRMWARE) from the controller
3.) Remove the four M4 nuts and change the board
4.) DIL-Switch to be set according to old controller. Re-fit EPROM and cables
5.) Note scan counter of new controller and, if at hand, load parameter file
6.) If parameter is not available, load default values and set values according to the values given
by InoTec
7.) For scanners of the type 2600 the sample delay has to be set and a white calibration
performed (see point 3.2.1.4)
8.) For scanners of the type 4000 perform a white calibration
Return the faulty controller board to InoTec together with the scanner’s serial number and scan counter
values.
.
Содержание SCAMAX 2600
Страница 7: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 7...
Страница 12: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 12...
Страница 14: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 14...
Страница 16: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 16 Illustration of optical reduction...
Страница 34: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 34...
Страница 49: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 49...
Страница 56: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 56...
Страница 62: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 62...
Страница 66: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 66...
Страница 67: ...SCAMAX 2600 4000 Tech Manual 06 2002 T 67...