NV200 Family User Manual
<< Back to Contents
Copyright © Innovative Technology Ltd 2016
Doc: NV200 Family Manual
Version: 1.0
Page 72 of 119
Click ‘Begin Testing’, this will initiate a scan of the unit to detect the appropriate
settings. Once the unit type has been detected the buttons will be populated, click
on the appropriate mechanism which needs to be tested.
9.5.3
Re-initialisation of the sensors
Should the NV200 be displaying unit not initialised (as explained in
Section 8.1
), or
the unit is showing some degredation in acceptance which the cleaning routines
outlined in
Section 7.2
then it may be necessary to re-initialise the sensors on the
NV200, this is possible through diagnostic tools.
N
avigate to the Initialise tab, click ‘Initialise’ the motors will activate, when you hear
the drive motor running continuously then insert the green calibration card. You will
now see the sensors calibrate. When complete a prompt will tell you the sensors had
been calibrated. Should you have any sensors which cannot calibrate contact your
local repair centre, a list of which can be found on our website.
Содержание NV200 Spectral
Страница 1: ...Doc GA02031 Version 1 0 ...