The functional sequences displayed here are the basis of INGUN product design. These can differ
depending on the test application.
5
A
DDITIONAL
F
UNCTIONS
MA
XXXX
Different versions of the manual test fixtures are available with additional functions as bellow.
The “note for usage” generally must be considered because not all optional functions/additional
functions can be combined and other aspects must be considered.
5.1
Protection conductor wiring and contact protection for dangerous volt-
ages
MA 2x09
MA xx11
MA xx12
MA xx13
MA xx13T
MA xx14
40509
39851
39852
39853
39855
39854
ATS MA 09
ATS MA 11
ATS MA 12
ATS MA 13
ATS MA 11
ATS MA 14
41809
41811
41812
41813
41811
41814
When testing with a dangerous voltage (>25V AC and >60V DC) the test fixture must be
equipped with protective equipotential bonding, and protection type IP3x according to DIN EN
60529 must be fulfilled. The customising according to the low-voltage directive (NSRL) consists
of the safety protection conductor wiring (yellow/green cable) to all relevant parts and the check
via insulation resistor at pre-defined places (limit 0.3
).
W
ARNING
:
In case of any changes of the protection conductor wiring (e.g. exchange of wires
or replacement of the cable gland), new measurement of the protection conductor
resistance is necessary!
Содержание MA 2x09 Series
Страница 1: ...Technical manual Manual Test Fixture MA xxxx with ATS MAxx...
Страница 55: ...Technical manual MA xxxx with ATS MAxx INGUN Subject to technical changes without notice E OE 54...
Страница 57: ...Technical manual MA xxxx with ATS MAxx INGUN Subject to technical changes without notice E OE 56...
Страница 58: ...Technical manual MA xxxx with ATS MAxx 57 INGUN Subject to technical changes without notice E OE...