Page 15
4. MAINTENANCE
4.1 PERIODIC MAINTENANCE
1. Periodic (1 per month) replacement or cleaning of filters on side panel.
2. Removal of debris buildup in laser channel using brush or vacuum. DO NOT use a
blower of any type.
3. Galvonmeter mirror cleaning.
4. Final lens Cleaning - White vinegar
5. Check Heat Exchanger fluid level.
4.2 LASER OPTICS CLEANING AND ALIGNMENT
Laser optics alignment should be performed by a qualified technician.
Print test Tag by pressing the print button and holding it in until print starts and then
release it. At the bottom edge of the test tag will be a thin line.
Adjust galvo mirror until the line is at the edge of the tag. Tighten the mirror clamp
then check for alignment again. Clean all optics with white vinegar and return all
covers.
4.3 CLEANING OPTICS
Gently wipe the surface of each optic with a coth that will not leave any fibers and is
not abrasive. The cleaning solution that is recommended is
white vinegar
.
DO NOT
RUB HARD.
Drag the cotton across the surface just fast enough so the liquid
evaporates right behind the cotton. This should leave no streaks. For optics that are
severely contaminated call InfoSight for replacement. After cleaning replace top
cover.
Do not used compressed air to clean optics!