Multifors 2 - Parallel Bench-top Bioreactor
Interferences
Page
200
of
211
28 March 2018
10.3 Interferences Basic Unit and Operating Panel
Interference
Equipment does not work. Power indicator of the main switch is not illuminated; monitor of the operating
unit remains dark.
Possible cause
Remedy
By
Power cable not connected.
Connect the power cable.
Operator
Main switch not switched on.
Switch on the main switch.
Operator
Fuses blown
Check fuses of mains connection.
Check fuses of basic unit.
Qualified elec-
trician
Break in power connections.
Check power plug.
Check power cable on damages or kinks.
Check socket and cable on basic unit.
Qualified elec-
trician
Interference
Power indicator of the main switch is illuminated; monitor of the operating unit remains dark.
Possible cause
Remedy
By
Monitor of operating unit is switched
off.
Switch the monitor of the operating unit on at the
ON/OFF key.
Operator
Power supply cable not connected to
the operating unit.
Plug the power supply cable in at the DC socket on
the rear side of the operating unit
Operator
Interference
No communication between equipment and operating unit.
Possible cause
Remedy
By
iDDC-bus cable (display cable) not
connected to the operating unit.
Plug the round connector of the iDDC-bus cable in
at the COM1 socket on the rear side of the operating
unit.
Plug the flat connector of the iDDC-bus cable in one
of the two sockets on the rear side of the basic unit.
Operator