Checking the system with the 

EASYGLUCO Auto-coding

TM

control solution 

The 

EASYGLUCO Auto-coding

TM

Control Solution is used to check that the meter

and the test strips are properly working together to provide you with the most

accurate reading.  

Note : In order to perform the more accurate glucose control solution test, you can use 

the control solution 3 different levels – low, normal and high.

You can buy each 

EASYGLUCO Auto-coding

TM

control solution (low, normal and high

levels) separately, please consult your local distributor about control solutions.

Control Solution Test can only be used with the 

EASYGLUCO Auto-coding

TM

System

and should be used during the following cases:

ҟ

When a new vial of test strips is opened.

ҟ

Any suspicion that the meter or test strips are not working properly.

ҟ

When your blood glucose test results are not consistent with your symptoms, or if you

think they are not accurate. 

ҟ

If you drop the meter.

ҟ

Use routinely to obtain accurate results.

When the control solution is applied to the top edge of the 

EASYGLUCO Auto-coding

TM

Test Strip, you should get results within the expected range printed on the label of the

test strip vial.  If the control solution test results fall outside the range, repeat the test.

Results that fall out-side the range may be caused by:

ҟ

Error in performing the test.

ҟ

Failure to shake the control solution vial well enough (must shake vigorously).

ҟ

Failure to discard the first drop of control solution.

CAUTION

1. If your glucose control solution falls out of range, do not perform a glucose test 

    and please contact Infopia customer support. (www.infopia21.com)

2. The glucose control solution range is not the recommended range for your blood glucose level.

3. Discard the used control solution and test strips carefully. Contactyour 

    Healthcare Provider forinstructions on disposal.

ҟ

Expired or contaminated control solution.

ҟ

Test Strip deterioration.

ҟ

Meter malfunction.

ҟ

Control solution that is too warm or cool.

16

17

Automatic code recognition

Caution

1. Do not bend the test strip to prevent the Automatic code recognition failure.

2. If the code recognition label is damaged, the code recognition may fail. Please

check the code number on the LCD display window with the code number on the

test strip vial.

1. 

EASYGLUCO  Auto-coding

TM

has

automatic code recognition function. 

2. It recognizes automatically the code

number. This is a very convenient

function preventing the inconvenience of

setting the code number on the meter

every time.

the                                            control solution

Checking the system with

Содержание EASYGLUCO Auto-coding

Страница 1: ...nfopiaCo Ltd 132 Anyangcheondong ro Dongan gu Anyang si Gyeonggi do Korea 431 836 http www infopia21 com ObelisS A Bd GeneralWahis531030Brussels Belgium Tel 32 2 732 59 54Fax 32 2 732 60 03 E mail mai...

Страница 2: ...Blood Glucose Monitoring System for selftesting...

Страница 3: ...O Auto codingTM System is for in vitro diagnostic use only A warranty registration card is included with your system Please complete the three year warranty card and send it to us This will ensure tha...

Страница 4: ...Auto codingTM Blood Glucose Monitoring System KIT Contents 8 EASYGLUCO Auto codingTM 9 r e t e M e s o c u l G d o o l B EASYGLUCO Auto codingTM Blood Glucose Meter Display 10 EASYGLUCO Auto codingTM...

Страница 5: ...Month and Year 2 Test strip ejector Your Blood Glucose Monitoring System has been sealed and should contain all of the parts shown above Please return the system to the place of purchase or contact yo...

Страница 6: ...YGLUCO Auto codingTM Test Strip vials in a cool dry place between 2 30 C Do not freeze 3 Make a notation of the discard date on the vial label when you first open it Discard remaining EASYGLUCO Auto c...

Страница 7: ...10 000 feet will have an effect on test results Doctors Please note the following interferences that may affect test results test strip results can be used with hematocrit levels in the range of 20 t...

Страница 8: ...ter re setting the test unit See also conversion table on page 38 2 Without setting the date properly the average glucose level and the results in memory will not show proper values It is recommended...

Страница 9: ...he test strip vial If the control solution test results fall outside the range repeat the test Results that fall out side the range may be caused by Error in performing the test Failure to shake the c...

Страница 10: ...ontrol Solution test results continuously falls outside the range printed on the vial the EASYGLUCO Auto codingTM system may not be functioning properly DO NOT use the system to test your blood until...

Страница 11: ...cet in a biohazard garbage container Sterile lancets is used with a spring operated lancing device COMPONENTS Depth Indicator Adjustable Tip Lancet Holder Clear Tip Ejection System Sliding Barrel Rele...

Страница 12: ...direction of arrow on test strip arrow up Please do not insert the test strip upside down When you insert the test strip into the meter the power automatically turns on with the code and temperature...

Страница 13: ...d the meter turns off automatically Testing Procedure Caution 1 If the countdown does not start do not add more blood to the test strip Discard the test strip an restart testing 2 If you do not conduc...

Страница 14: ...setup press and button to select the correct User ID then press caution 1 If you did not activate the User Activity option and or the User ID option Step 4 1 and or Step 4 2 procedures are not conduct...

Страница 15: ...alues can be recalled in the LCD display at any time The newest result will replace the oldest beyond the 365 data stored Press shortly If the User ID option has been disabled during setup the newest...

Страница 16: ...er significantly from fingertip samples The conditions in which these differences are more likely to occur are when your blood glucose is changing rapidly such as following a meal as insulin dose or a...

Страница 17: ...d sample is taken If there is not enough blood gently massage the area until a sufficient sample has been collected caution To receive accurate test results at least 1 0 minimum sample volume is requi...

Страница 18: ...o not refrigerate Avoid exposure to sun 4 Close the vial cap tightly immediately after removing a EASYGLUCO Auto codingTM strip Please refer to your EASYGLUCO Auto codingTM strip instruction if you wa...

Страница 19: ...codingTM meter uses two 3V Lithium battery CR2032 The batteries are included in the set box When replacing the batteries only CR2032 or equivalent lithium batteries can be used Before replacing the b...

Страница 20: ...ccuracy of your test strip by performing a control solution test After re test with blood sample two or three times When HI message persists please consult a doctor for assistance Problem with the tes...

Страница 21: ...electrical current that will stimulate a chemical reaction This reaction is measured by the EASYGLUCO Auto codingTM Meter and displayed as your blood glucose result Note Different levels of reactions...

Страница 22: ...20 0 250 260 270 288 300 306 325 342 360 20 8 22 2 23 0 24 0 25 0 26 4 27 7 30 0 33 3 375 400 414 432 450 475 500 540 600 Sample type Sample volume Test Range Reading time Calibration Operating Tempe...

Страница 23: ...in the European community Three Year Warranty If at any time during the first three years after purchasem the meter does not work for any reason with the exception of obvious abuse misuse or disastrou...

Страница 24: ...46 47 Memo Memo...

Страница 25: ...48 Memo IMC00630 Rev 2014 05 03 INFM16B...

Отзывы: