Testing Your Blood - Preparation

Testing procedure 

1. Prior to use, shake the bottle gently several times.

2. Invert bottle and discard the first drop of control solution.

3. Wipe the dispenser tip clean.

4. Test strip is inserted fully into the test slot of the 

EASYGLUCO Auto-

coding

TM

meter.

5. Be sure the code number appearing on the LCD window matches the test

strip code on the strip vial.

6. When the strip sign blinks, press

Ƈ

or 

Ɖ

button once. The symbol 

<s 

for

control solution measuring process appears. Effect: The result is marked.

7. Hold the control vial at an angle with the tip near the end of the test strip.

Gently squeeze the vial and form a small drop. Pick up the meter and touch

the end of the test strip to the hanging drop. Be sure to hold the test strip to

the drop until the “beep” sounds.

8. After the “beep” sound, wipe the rest of the drop with tissue or cotton gauze.

If the drop is taken back into the control vial, it is possible to contaminate the

whole vial of control solution!

9. The displayed test result should be within the range printed on the test strip

vial.

Second application tip : Place a drop on clean aluminum foil, glass or plastic plate and

touch the end of the test strip to that drop. This method is more favorable, in preventing

contamination to the whole vial of control solution

IMPORTANT NOTE : If the 

EASYGLUCO Auto-coding

TM

Control Solution test results,

continuously, falls outside the range printed on the vial, the 

EASYGLUCO Auto-

coding

TM

system may not be functioning properly.  DO NOT use the system to test your

blood until you get a control solution test result that falls within the range. If you continue

to have problems, call your Customer Support and Service.

In the User ID option is activated, the control solution test result is stored, but can only be

recalled after deactivating the User ID option.

Before testing, be sure to read this section and test strip package insert found

in the test strip box carefully. Make sure you have all the necessary items to

begin the test:

1. 

EASYGLUCO Auto-coding

TM

Meter

2. 

EASYGLUCO Auto-coding

TM

Test

Strip

3. Lancing Device

4. Sterile lancets

Caution 

To reduce risk of infection :

1. The lancing device and sterile lancets should NOT be shared with others.

2. Always use a new, sterile lancet and a new strip

3. Practice using the Lancing Device and become accustomed with its use.

4. Lancets and strips are for single use only.

5. Wash your hands with warm, clean water and soap before testing.

6. Avoid getting hand lotion, oils, dirt or debris on the lancets or on the lancing

device.

7. Dry your hands completely before testing.

8. Sterile lancets are for single use only. Do not reuse. 

9. Sterile lancets is used with a spring-operated lancing device

10. Sterile lancets sterilized using gamma irradiation. After opening, exposed parts   

should not be touched or put on the floor to prevent contamination.

the                                            control solution

Checking the system with

Содержание EASYGLUCO Auto-coding

Страница 1: ...nfopiaCo Ltd 132 Anyangcheondong ro Dongan gu Anyang si Gyeonggi do Korea 431 836 http www infopia21 com ObelisS A Bd GeneralWahis531030Brussels Belgium Tel 32 2 732 59 54Fax 32 2 732 60 03 E mail mai...

Страница 2: ...Blood Glucose Monitoring System for selftesting...

Страница 3: ...O Auto codingTM System is for in vitro diagnostic use only A warranty registration card is included with your system Please complete the three year warranty card and send it to us This will ensure tha...

Страница 4: ...Auto codingTM Blood Glucose Monitoring System KIT Contents 8 EASYGLUCO Auto codingTM 9 r e t e M e s o c u l G d o o l B EASYGLUCO Auto codingTM Blood Glucose Meter Display 10 EASYGLUCO Auto codingTM...

Страница 5: ...Month and Year 2 Test strip ejector Your Blood Glucose Monitoring System has been sealed and should contain all of the parts shown above Please return the system to the place of purchase or contact yo...

Страница 6: ...YGLUCO Auto codingTM Test Strip vials in a cool dry place between 2 30 C Do not freeze 3 Make a notation of the discard date on the vial label when you first open it Discard remaining EASYGLUCO Auto c...

Страница 7: ...10 000 feet will have an effect on test results Doctors Please note the following interferences that may affect test results test strip results can be used with hematocrit levels in the range of 20 t...

Страница 8: ...ter re setting the test unit See also conversion table on page 38 2 Without setting the date properly the average glucose level and the results in memory will not show proper values It is recommended...

Страница 9: ...he test strip vial If the control solution test results fall outside the range repeat the test Results that fall out side the range may be caused by Error in performing the test Failure to shake the c...

Страница 10: ...ontrol Solution test results continuously falls outside the range printed on the vial the EASYGLUCO Auto codingTM system may not be functioning properly DO NOT use the system to test your blood until...

Страница 11: ...cet in a biohazard garbage container Sterile lancets is used with a spring operated lancing device COMPONENTS Depth Indicator Adjustable Tip Lancet Holder Clear Tip Ejection System Sliding Barrel Rele...

Страница 12: ...direction of arrow on test strip arrow up Please do not insert the test strip upside down When you insert the test strip into the meter the power automatically turns on with the code and temperature...

Страница 13: ...d the meter turns off automatically Testing Procedure Caution 1 If the countdown does not start do not add more blood to the test strip Discard the test strip an restart testing 2 If you do not conduc...

Страница 14: ...setup press and button to select the correct User ID then press caution 1 If you did not activate the User Activity option and or the User ID option Step 4 1 and or Step 4 2 procedures are not conduct...

Страница 15: ...alues can be recalled in the LCD display at any time The newest result will replace the oldest beyond the 365 data stored Press shortly If the User ID option has been disabled during setup the newest...

Страница 16: ...er significantly from fingertip samples The conditions in which these differences are more likely to occur are when your blood glucose is changing rapidly such as following a meal as insulin dose or a...

Страница 17: ...d sample is taken If there is not enough blood gently massage the area until a sufficient sample has been collected caution To receive accurate test results at least 1 0 minimum sample volume is requi...

Страница 18: ...o not refrigerate Avoid exposure to sun 4 Close the vial cap tightly immediately after removing a EASYGLUCO Auto codingTM strip Please refer to your EASYGLUCO Auto codingTM strip instruction if you wa...

Страница 19: ...codingTM meter uses two 3V Lithium battery CR2032 The batteries are included in the set box When replacing the batteries only CR2032 or equivalent lithium batteries can be used Before replacing the b...

Страница 20: ...ccuracy of your test strip by performing a control solution test After re test with blood sample two or three times When HI message persists please consult a doctor for assistance Problem with the tes...

Страница 21: ...electrical current that will stimulate a chemical reaction This reaction is measured by the EASYGLUCO Auto codingTM Meter and displayed as your blood glucose result Note Different levels of reactions...

Страница 22: ...20 0 250 260 270 288 300 306 325 342 360 20 8 22 2 23 0 24 0 25 0 26 4 27 7 30 0 33 3 375 400 414 432 450 475 500 540 600 Sample type Sample volume Test Range Reading time Calibration Operating Tempe...

Страница 23: ...in the European community Three Year Warranty If at any time during the first three years after purchasem the meter does not work for any reason with the exception of obvious abuse misuse or disastrou...

Страница 24: ...46 47 Memo Memo...

Страница 25: ...48 Memo IMC00630 Rev 2014 05 03 INFM16B...

Отзывы: