Manual of OMNI Express Patient Monitor
82
037.007.0001 Rev.2
TRENDING INTERVAL
Trending interval denotes how often the Graphic trend or the Tabular Trend displays the
data.
There are eight items for trending interval choosing: 5s, 10s, 15s, 30s, 1 min, 5mins,
10mins and 15mins.
For instance, if 5s is chosen as the reference trending interval, then we can recall the
trend data displayed in the trend every 5s.
TREND GRAPH ANALYSIS
TREND GRAPH ADMITTANCE
Press the
“Trend Graph” button to open the Graphical Trend window.
Figure 29: Window for Trend Graph
Each page displays the trend chart for one parameter, and the user can change the graph
by choosing
PARAMETER
. The parameter order is as follows:
HR, SpO
2
, RESP, NIBP, ST, PR, TEMP, EtCO
2
an so on.
The newest data is on the right side of the graph, time is displayed on the bottom of the
graph on a 24 hour scale, and the upper and lower limits of the parameter are displayed
on the left side of graph.
CURSOR BAR
The cursor bar is the red vertical line on the trend graph. The parameters
’ values in the
graph are obtained at the time that the red vertical line indicates.
Press the
“Shift Left” or “Shift Right” button. You will move the red cursor bar left or right
until it is at the appropriate position.
TABULAR TREND ANALYSIS
TREND TABLE ADMITTANCE
Press the
“Tabular Trend” button to open the Tabular Trend window. The window will
display in the waveform area on the screen.
Sixteen groups of parameters are listed on every page with three hundred groups in total.
These data will be listed in order from newest to oldest. Time is displayed in 24 hour
format. The parameter name is display on the top of chart and the invalid data will not
display.
BASIC TABULAR TREND