User’s
Manual for OMNI Express Patient Monitor
107
037.007.0001 Rev.2
Immunity test
IEC
60601
test level
Compliance
level
Electromagnetic
environment
–
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Only ISA CO2 is
tested at 20 V/m
3 Vrms
150 kHz to 80
MHz
3 V/m
80%AM@2H
z
80 MHz to
2.5 GHz
20 V/m
80%AM@1k
Hz
80 MHz to
2.5 GHz
3 Vrms
3 V/m
20 V/m
Portable
and
mobile
RF
communications equipment should
be used no closer to any part of the
OMNI Express Patient Monitor,
including
cables,
than
the
recommended separation distance,
which can be calculated using the
formula applicable to the frequency
of the transmitter.
Recommended separation distance:
P
d
2
.
1
=
80 MHz to 800 MHz
P
d
2
.
1
=
P
d
3
.
2
=
800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts
(W)
according
to
the
transmitter manufacturer and d is
the
recommended
separation
distance in meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an
electromagnetic
site
survey
a)
,
should be less than the compliance
level in each frequency range
b)
.
Interference may occur in the vicinity
of equipment marked with the
following symbol: