Manual of OMNI Express Patient Monitor
- 59
037.007.0001 Rev.2
MAINTENANCE AND CLEANING
SCHEDULE
The CO
2
Module flow rate accuracy should be verified by direct measurement using a
calibrated flow meter every 12 months.
CLEANING
Cleaning the CO
2
Module case, cable and connector:
1. Use a cloth dampened with isopropyl alcohol 70%, a 10% aqueous solution of sodium
hypochlorite (bleach), a 2% gluteraldehyde solution, ammonia, mild soap or disinfectant
spray cleaner such as Steris Coverage® Spray HB.
2. Wipe down with a clean water-dampened cloth to rinse and dry before use. Make
certain that the sensor windows are clean and dry before reuse.
Cleaning the Sidestream on-Airway Adapters and Sidestream Sampling Kits:
Sidestream on-airway adapters and sidestream sampling kits are for single patient use.
Treat in accordance with hospital protocols for handling single patient use devices.
[NOTE]
DO NOT submerge or sterilize the CO
2
Module.