3 - 16
XTC/3 Operating Manual
1
Make sure XTC/3 is in
READY
or
IDLE
.
2
Press
START
. The next layer to be run will enter pre-deposition, and continue
on through deposition and post-deposition.
3
When the layer is complete, XTC/3 will go to
IDLE
. If
AutoStart Next Layer
(XTC/3M only) has been set to
Yes
in the
Process
sub-screen of the
General
Parameter
display, the next layer, if there is one, will be started automatically.
If not, press
START
again to begin the next layer. Repeat until the process is
complete.
4
If at some point there is a need to interrupt or discontinue the process, press
STOP
. This will close the sensor and source shutters, set the control voltage
output to zero, and freeze the display. The process can be restarted where it
left off by pressing
START
. (The pre-deposition phases will be repeated.) To
completely abandon the run, press
RESET
after
STOP
.
5
A critical error may occur while a process is running.
For example, a crystal failure on a single crystal sensor head may occur in
pre-deposition. XTC/3 will automatically
STOP
or go into
Time-Power
(if the
crystal failure occurs while in
Deposit
and the
Time-Power
option is set to
Yes
). Assuming the error has been remedied, the process can be continued
where it left off by pressing
START
. Pressing
RESET
will abandon the run.