6 - 2
XTC/3 Operating Manual
DIGITAL INPUT
Unit has gone into
STOP
because the designated Stop Input was activated.
EMPTY PROCESS
Indicates that the selected process has no layers and therefore cannot
be selected to run.
END OF PROCESS
Indicates that the last layer in the process has reached the Idle state.
ETCH MODE
Indicates that XTC/3 is in the etch mode.
ETHERNET IN USE
Ethernet values cannot be changed at this time.
ETHERNET VALUES HAVE CHANGED
Changes take effect on power up.
FRONT PANEL
Indicates that XTC/3 was stopped by pressing
STOP
on the front panel.
HANDCONTROL
Indicates that XTC/3 was stopped by pressing
STOP
on the
Handheld Controller.
INCORRECT TIME
Incorrect value was entered.
INPUT INCORRECT
Incorrect lock code was entered to unlock the unit.
INVALID MASK
Gateway mask value is Incorrect.
L LOCK
Indicates XTC/3 is in a Local Program Lock mode. This prohibits any
parameter from being entered from the front panel until the Program Lock
Code is entered.
LAYER INSERT
Appears in Process List to indicate that Layer Insert was successful.
MAX POWER
Indicates the specified maximum power has been reached. The power value
will be displayed in red.
NO BACKUP CRYSTAL
An attempt is being made to start a layer and there are no good crystals
remaining in the sensor head. This message will appear unless ’Start without
Backup’ is enabled.
NO GOOD XTAL TO SW
There are no good crystals available to switch to.