INFICON
Warning and error messages | 7
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
97 / 152
Type
Notification
Possible sources of
error
Remedy
W110
Real-time clock was reset!
Enter date and time
The real-time clock has
not been set
• Enter the correct date and time
• Check that the message does not appear
when you restart the device
Battery jumper on the
MSB not set
• Contact customer service
Battery on MSB is
discharged or defective
• Contact customer service
Real-time clock
defective
• Contact customer service
W111
Many EEPROM write cycles
within the last 6 minutes
A very large number of
write commands were
issued through
interfaces. Over time,
this reduces the service
life of the EEPROM.
• Remove the excess write accesses from
your control program
E112
Impermissible DIP switch
setting
The DIP switch on the
MSB has an invalid
position
• Contact customer service
MSB defective
• Contact customer service
W113
The device ID does not
match the recognized
backing pump
The DIP switch on the
MSB has an invalid
position
• Contact customer service
Type of backing pump
was not or was
incorrectly detected
• Contact customer service
MSB defective
• Contact customer service
W122
No response from bus
module
Connection to BUS
module interrupted
• Check the connection to the bus module
• Replace the connection cable to the bus
module
Bus module defective
• Replace the bus module
Bus module connection
on the device defective
• Contact customer service
Содержание UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 Factory setting 70 ...
Страница 151: ......
Страница 152: ......