5 | Installation
INFICON
50 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
6
In case you would like to suppress any possible existing background signals
(helium background in the test object) press the ZERObutton.
In order to reverse the background suppression press the ZERO button on the
control panel for 2 - 3 seconds, see "ZERO button [
25]”.
7
Press the Stopbutton.
ð
The device switches to the "Standby" mode.
If you press the Stop button on the control panel for a few seconds the inlet of
the device is ventilated.
8
If you want to end the test now you can switch off the device.
9
If you want to check the internal calibration wait, for getting a better quality of
measurement results, 15 to 20 minutes until the device has warmed up.
10
To enter the calibration menu, press the
button.
11
Select “Internal“.
12
Press the
button.
ð
The automatic internal calibration starts and requires approx. 30 seconds.
13
If you want to check the measurement accuracy of the device using the optional
helium calibration leak remove the blank flange from the inlet and connect an
open helium calibration leak onto the inlet.
14
Press the STARTbutton.
ð
The inlet is evacuated and the leak rate of the test object is measured and
displayed.
15
Press the Stopbutton to interrupt the measurement.
ð
The leak detector changes into the "Standby" mode.
16
Press the Stop button on the housing until the message STANDBY/VENTED
appears on the display.
ð
The inlet is now in the vented state.
17
Separate the helium calibration leak from the inlet and flange the inlet blind
again.
18
Switch off the device via the power supply switch.
Содержание UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 Factory setting 70 ...
Страница 151: ......
Страница 152: ......