To the Kashiyama NeoDry backing pump
When pumping clean gases such as dry air and inert gases, the recommended
schedule for preventive maintenance is 3 years.
When pumping condensable gases such as water vapour and gas-mixed solvents as
well as aggressive or corrosive gases, it may be necessary to reduce the maintenance
interval to one year.
The recommendation is made to maintain the technical specifications of the pump
specified by the manufacturer throughout the life of the device. Lack of maintenance
will not normally result in a sudden failure of the pump. Due to wear, pressure
performance will decrease over time.
As long as the pressure performance is suitable for the customer application, the
replacement interval can be changed according to the customer's request.
About devices with I•RISE
Devices with built-in I•RISE function do not require any additional maintenance under
clean working conditions.
Содержание UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 01 2206 Factory setting 70 ...
Страница 151: ......
Страница 152: ......