Transpector MPS Operating Manual
3.5 The Ion Detector
The ion detector region of the sensor consists of the quadrupole exit lens and the
detector itself. Often, the quadrupole exit aperture is biased negatively with respect
to the anode, focusing ions that have been transmitted through the quadrupole into
the detector element. The detector can be a simple Faraday Cup (FC), an Electron
Multiplier (EM), or a combination of both.
3.5.1 The Faraday Cup Detector
The Faraday Cup detector is typically a metal plate or a cup-shaped electrode, on
which the ion beam impinges. Ions strike the detector and are neutralized, thus
drawing a current from the circuitry connected to the electrode. Usually, the current
flow that results is exactly equal to the incident ion current. In Transpector MPS
instruments, the Faraday Cup is at ground potential.
The sensitivity for Transpector MPS instruments equipped with a simple Faraday
Cup detector is typically 3 x 10
-4
amps per Torr. The detected currents can be as
small as 1 x 10
-15
amps for ultra-high vacuum levels. See
.
Figure 3-5 Faraday Cup detector
Detector
Quadrupole Exit Lens
Faraday Cup
Signal Output