6 - 6
Transpector MPH Operating Manual
NOTE:
Although the following measurements may measure below 30 M
with the
filament assembly and ceramic shield in place, they must be above 30 M
when measured with the filament assembly and ceramic shield removed.
2
Measure the resistance of each of the pins with respect to ground
(pin
7
and
8
). These measurements must be above 30 M
.
3
Measure the resistance of each of the pins with respect to each other. All of
these measurements must be above 30 M
, with the exception of the
measurements between each filament pin and the filament common pin if the
filaments have not failed.
6.5.3 Filament Kit Replacement
A filament replacement kit can be purchased from INFICON. This kit contains a
new filament assembly mounted on a shipping fixture and a small Allen wrench.
Perform the following steps to replace the filament.
NOTE:
Refer to
section 6.5.1.1, Tools for Replacing the Filament Kit, on page 6-4
before continuing.
CAUTION
Neither the filament assembly or the ion source can be
cleaned. When dirty, they must be replaced.
1
Using the 3 mm nut driver, remove the three nuts and lock washers (
A
)
from the standoffs.
Figure 6-2 Removing the nuts
A