48
tina56e1-f (2015-09)
The current ATM threshold setting
•
can be read / written via the diagnostic port
•
is output at the measurement signal output instead of the
pressure signal, can be measured with a voltmeter, and is
displayed on the LCD display after the <SP1> or <SP2>
button is pressed
•
can be read / written via the DeviceNet, Profibus and RS485
interface.
DANGER
DANGER: malfunction
If processes are controlled via the signal output,
keep in mind that by pushing the <SP> button the
measurement signal is suppressed and the cor-
responding threshold value is output instead. This
can cause malfunctions.
Push the <SP> button only if you are sure that no
malfunction will cause.
Programming ATM setpoint
Programmable parameters:
(
→
[3], [4], [5], [6])
Factor of ATM
Low Trip Enable
Low Trip Point Hysteresis
High
Trip
Enable
High Trip Point Hysteresis
Setpoint
Mode
Switching characteristics of the setpoints
→
41.
Diagnostic port
→
[3].