40
tina83e1-d (2018-12)
For adjusting the zero, operate the gauge under the same con-
stant ambient conditions and in the same mounting orientation
as normally.
The zero can be adjusted via
•
the <ADJ> button on the gauge,
•
the diagnostic port (
→
•
the RS232C/485C interface (
→
•
the EtherCAT interface (
→
If you are using a seal with centering ring and filter, check
that they are clean or replace them if necessary
(
→
"Deinstallation").
Put the gauge into operation and operate it at atmospheric
pressure for at least 10 minutes.
Briefly press the <ADJ> button with a pin (max. ø1.1 mm)
and the ATM adjustment is carried out, or …
ATM
HV
… perform the adjustment via the diagnostic port or via the
EtherCAT interface.
The Pirani sensor is adjusted to 1000 mbar (duration
≈
5 s).