Remove one (1)
eXact
®
Strip Micro
, part number is dependent upon the test being run,
from the bottle before beginning the test. Set the strip in a dry, convenient place and
recap the bottle immediately.
REMOVE STRIP
6
Dip the
eXact
®
Strip Micro (or add Reagent)
into the
CELL
immediately press
READ
. This starts
the
20 SECOND
countdown timer. During this time move the strip in a gentle back and forth
motion.
Remove and discard the strip when the time displays 1
. The cursor will move across
the display, informing you that it is about to measure the sample. Record result displayed (this
result is automatically stored in TR7). After testing is completed, rinse cell immediately.
DIP STRIP AND PRESS “READ”
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
100 %T
.
The sample is ready for testing.
ZERO METER
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the
MENU
button until the display shows the parameter TR7.
Tests listed on pages 10 through 26 will require the TR7 MENU and the conversion charts
included. Other tests listed on page 31 will require the use of a conversion chart available at
www.sensafe.com.
SELECT TEST: TR7
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing
minimizes the potential for cross-contamination from a previous test. Finally, fill cell to
capacity (4ml) with the water sample.
FILL METER WITH SAMPLE
Transmission Test Procedure
TR7: Different ions react with a specific indicator to form a color or a precipitate that proportionally indicates
the concentration of the ion present by the transmission value found. A conversion chart is then used to
determine the ion concentration using the transmission value. The advantage of using transmission
measurement is that many different ions can be determined in one MENU, which expands the flexibility of
this meter. Once you have determined the %T result for the test you ran, find this %T result in the conversion
chart and read the concentration corresponding for this %T. The Micro 7+ gives the %T values as 3 digits
(example 99.2) but only the first two digits are used in the charts (round off your %T value to two digits). Tests
that can be used with the Micro 7+ in TR7 MENU are listed on page 31. Typically, the most common tests are
supplied with the conversion chart in this booklet. For conversion charts and information about the tests not
given in this booklet, visit our website. You can also use the TR7 MENU to develop your own custom
conversion chart for unusual samples to get more accurate results.
TR
TR7
Acid pH Table
Alkali pH results require the table below. Follow
eXact
®
Micro 7+ Acid pH Test
Procedure
(above) using
eXact
®
Strip Micro Acid pH, Part No. 486624
eXact
®
Strip Micro Acid pH, Part No. 486624 - for 4mL Samples
Rev. 020609 AcpH
tr
90
80
70
60
50
40
30
20
10
0
9
8
7
6
5
4
3
2
1
0
<3.0
3.3
4.2
4.7
5.1
5.4
5.8
6.2
>6.2
>6.2
<3.0
3.4
4.2
4.7
5.1
5.5
5.8
>6.2
>6.2
>6.2
5
Dip the
eXact
®
Strip Micro Acid pH, Part No. 486624
into the
CELL
and immediately press
READ
. This starts the
20 SECOND
countdown timer. During this time move the strip in a
gentle back and forth motion.
Remove and discard the strip when the time displays 1
. The
cursor will move across the display, informing you that it is about to measure the sample.
Record result displayed (this result is automatically stored in TR7).
DIP STRIP AND PRESS “READ”
6
Find the “TR” result in the table below to determine the pH. (Example: a “TR” result of 65.3
(use only the 65 for the chart) equals a pH value of 4.9). Record result. After testing is
completed, rinse cell immediately.
USE TABLE
4
Press the
ZERO/ON
button. The cursor will move across the display, followed by
100 %T
.
Sample is ready for testing.
ZERO METER
1
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
2
Press and re-press the
MENU
button until the display shows the parameter TR7.
SELECT TEST: TR7
3
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml) with
the water sample.
FILL METER WITH SAMPLE
Acid pH Test Procedure
<3.0
3.5
4.3
4.8
5.2
5.5
5.9
>6.2
>6.2
>6.2
<3.0
3.6
4.3
4.8
5.2
5.5
5.9
>6.2
>6.2
>6.2
<3.0
3.7
4.4
4.9
5.2
5.6
5.9
>6.2
>6.2
>6.2
<3.0
3.8
4.4
4.9
5.3
5.6
6.0
>6.2
>6.2
>6.2
3.0
3.9
4.5
4.9
5.3
5.7
6.0
>6.2
>6.2
>6.2
3.1
3.9
4.5
5.0
5.3
5.7
6.1
>6.2
>6.2
>6.2
3.2
4.0
4.6
5.0
5.4
5.7
6.1
>6.2
>6.2
>6.2
3.2
4.1
4.6
5.1
5.4
5.8
6.2
>6.2
>6.2
>6.2