Industrial Test Systems 486691 Скачать руководство пользователя страница 2

eXact

®

 Micro 7+ Meter Specifications

We offer a “Green” Alternative

2

Measurement Method: 

Photometric

Light Source:   

Light Emitting Diode (LED)

Wavelength:  

525 nm

Transmission Range: 

100 - 0.00 %T

Photometric Precision: 

+/- 0.1/0.01 %T

Automatic Range Selection: 

See Specifications below

Display:

 

 

3-digit customized liquid crystal display

  with 

annunciators

CELL Pathlength:

 20mm

Cell Chamber:   

    

Custom-molded, proprietary, PET plastic

 

 

    fused into chamber, non-removable

Sample Required: 

    

4 ml (0.13 oz)

Operating Temperature Range:    

0 - 50°C (32° - 122°F)

Power Supply:  

    

(4) AAA alkaline batteries

Battery Life: 

 

    

>2000 tests with alkaline batteries

Electromagnetic Compliance:     

Emitted Interference - EN 61326

(EMC)

 

 

    Immunity to Interference - EN 61326

Waterproof Rating:

 

    Exceeds IP67

Weight:

 

 

    Instrument: 140 g (5 oz)

Dimensions:

   

    Instrument: 5 (W) x 3.5 (D) x 16.5 (H) cm;

 

 

    (2 x 1.4 x 6.375 in)

eXact

®

 Micro 7+ Photometer

Menu Tests for 

Range 

Resolution 

+/- Accuracy

CL1 

Free Chlorine (DPD-1) 

0.00-2.39 ppm 

0.01 

0.02

 

& Total Chlorine (DPD-3)  2.4-6.0 ppm 

0.1 

0.1, or 4%

PH2 pH 

6.2-8.4 

pH 

0.1 

0.3

BR3 

Bromine (DPD-1) 

0.00-2.99 ppm 

0.01 

0.03

 

 

3.0 - 9.0 ppm 

0.1 

0.1, or 4%

AL4 Total 

Alkalinity 

20-180 

ppm  1 

25

CA5 

Calcium as CaCO

3

 

20-990 ppm 

10 

20 or 7%

CU6 Copper 

(Cu

+2

) 0.00-1.99 

ppm 

0.01 

0.02

 

 

2.0 - 8.0 ppm 

0.1 

0.1 or 4%

TR7 Transmission 

99.9-10.0 

%T 0.1 

0.1

 

(17 other test parameters) 

9.99-0.01 %T 

0.01 

0.01

eXact

®

 Micro 7+ Specifications

cro

eX

F

E

D

C

B

A

ter Specifications

G

H

+

ITEM PICTURED

LETTER

COMPONENT

NAME

A

Mixing Cap

B

Cell (Built-in Plastic, 4 ml)

C

LCD Display

D

READ

 Button

E

MENU

 Button

F

ZERO/ON

 Button

G

Battery Cover with loop for lanyard

H

AAA Batteries (x4)

eXact

® 

Strip Micro 7+ 

has been designed to offer the user a more “Green” and cost-effective alternative to testing.

Instead of using a 10ml water sample, 

eXact

® 

Strip Micro 7+ 

uses a 4ml water sample, which uses up to 60% less

chemical per test. The accuracy of the meter is maintained by designing the photo cell with a 20mm pathlength.

(Used for Cyanuric

Acid, Fluoride, Iron,

and Manganese tests)

31

* Results utilize the Tr-7 (Transmission) meter function and require the use of a conversion table. See respective test procedures
for more information and tables.

** Total Chlorine DPD-3 Test requires Free Chlorine DPD-1 (486637) to be run first.

Alizarin Red S + Citrate

DPD

Oxalic Acid

DPD

Kl

DPD + Kl

Biquinoline

DPD + Kl

DPD

Phenol Red

Salicylate Method

Silver (ppt)

DPD

Diphenylcarbazide

Melamine (ppt)

SPADNS

DPD

TPTZ

1,10 Phenanthroline

Phthalein Purple

PAN + Cyanide

Zinc Reduction

Chromotropic Acid

Alizarin Red S

Thymol Blue

Tetraphenylborate

Barium (ppt)

Turbidity values as NTU

PARAMETER

No.

PART
NUMBER

# OF

TESTS

DETECTION

RANGE

CHEMISTRY

eXact

®

 Strip Micro (4mL ) Reagent Specifications - For use with eXact

®

 Micro 7+, Part no. 486691

Alkalinity, Total

Bromine (DPD-1)

Calcium (as CaCO

3

)

Chlorine, Free (DPD-1)

Chlorine, Total (DPD-3)**

Chlorine, Total (DPD-4)

Copper (Cu

+2

)

Ozone (DPD-4)

Permanganate (DPD-1)

pH

Ammonia*

Chloride (as NaCl)*

1:40 Dilution of sample

Chlorine Dioxide (DPD-1)*

Chromate (Chromium)*

Cyanuric Acid*

Fluoride*

Iodine (DPD-1)*

Iron (Fe

+2

)*

Iron (Fe

+2

/Fe

+3

)*

LR Total Hardness (as CaCO

3

)*

Manganese*

Nitrate (as NO

3

)*

Nitrite (as NO

2

)*

pH, Acid*

pH, Alkali*

Potassium*

Sulfate*

Turbidity* (as transmission)

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

17

18

19

20

21

22

23

24

25

26

27

28

486641

486637

486629

486637

486638

486670

486632

486670

486637

486639

483343-M

481657

486637

486614

481652

486643

486637

486631

481623

486630

486606

486617

486623

486624

486609

486621

486608

None

100

100

50

100

100

100

50

100

100

100

25

25

100

50

130

25

100

50

50

100

24

50

50

50

50

50

50

No Reagent

20 - 180 ppm

0.01 - 9.0 ppm

20 - 990 ppm

0.01 - 6.0 ppm

0.01 - 6.0 ppm

0.01 - 6.0 ppm

0.04 - 8.0 ppm

0.01 - 10.0 ppm

0.01 - 4.5 ppm

6.2 - 8.4

0.03 - 5.8 ppm

3 - 285 ppm

100 - 11000 ppm

0.03 - 6.5 ppm

0.02 - 4.89 ppm

5 - 62 ppm

0.1 - 1.1 ppm

0.01 - 10 ppm

0.01 - 2.00

0.15 - 3.5 ppm

2 - 77 ppm

0.03 - 1.5 ppm

13 - 140 ppm

0.05 - 3.5 ppm

3.0 - 6.2

8.3 - 9.6

2 - 200 ppm

1 - 111 ppm

3 - 580 NTU

PRICES & SPECIFICATIONS SUBJECT TO CHANGE WITHOUT NOTICE.

If your required procedure is not listed in this manual, please see the back page 

for our contact information.

NOTE: Because most of our products are test strips or use reagents 
that have little or no hazard in the quantity sold, MSDS sheets are 
not supplied with the test. The exceptions are the Manganese 
(486606) test, which comes with 2 strips and one liquid reagent 
(PAN); Fluoride (486643) test, which is a liquid reagent (SPADNS), 
and Iron (481623) test, which is a powder reagent. 

To ensure optimal performance, store your eXact

®

 kit in a cool, dry place away from

excess heat (below 100°F / 38°C), moisture, and oxidizers such as Chlorine and Bromine.

eXact

®

 Strip Micro 7+ Reagent Reorder Information

Содержание 486691

Страница 1: ...3 329 9743 ITS SENSAFE COM WWW SENSAFE COM www poolcheckonline com For European Middle East Inquiries and Re Orders ITS Europe LTD Unit B Stanley Crt Glenmore Business Park Telford Rd Churchfields Industrial Estate Salisbury Wiltshire SP2 7GL United Kingdom Tel 44 0 1722 329502 Fax 44 0 1722 329880 SALES SENSAFE COM WWW ITSEUROPE CO UK www poolcheckonline com eXact Strip Micro DPD 1 Accuracy eXact...

Страница 2: ...7 to be run first Alizarin Red S Citrate DPD Oxalic Acid DPD Kl DPD Kl Biquinoline DPD Kl DPD Phenol Red Salicylate Method Silver ppt DPD Diphenylcarbazide Melamine ppt SPADNS DPD TPTZ 1 10 Phenanthroline Phthalein Purple PAN Cyanide Zinc Reduction Chromotropic Acid Alizarin Red S Thymol Blue Tetraphenylborate Barium ppt Turbidity values as NTU PARAMETER No PART NUMBER OF TESTS DETECTION RANGE CHE...

Страница 3: ...0 The following are some common messages that may be displayed including error messages If an error message other than those listed below is displayed please contact technical support in the USA at 803 329 0162 ext 0 LCD Message HI LO LO ER Description In READ mode test sample concentration is above the measurement range test specific In READ mode test sample concentration is below the measurement...

Страница 4: ... for pH and Total Alkalinity Sample water that may splash out of the CELL during movement of the eXact Strip Micro will not affect accuracy as long as the cell is 50 full at end of test Rinse the CELL with clean water immediately after completing the test Some test reagents will stain or coat the CELL Just before testing rinse the sample CELL with the sample water several times to get a representa...

Страница 5: ... time the KI is added For compliance testing you must time the two minutes and then make your measurement NOTE From testing in our lab water samples above 70 F 20 C generally reach a stabilized reading quickly eXact Strip Micro CL DPD 1 DPD 3 DPD 4 Interferences part nos 486637 486638 486670 Interfering Substance Acidity Alkalinity Bromine Bromamines Br2 Chlorine Dioxide ClO2 Copper Cu 2 Iodine I2...

Страница 6: ...y 0 01 press READ again and repeat After testing is completed rinse cell immediately Record the Total Chlorine as the highest value the meter displayed NOTE Standard Method 4500 Cl G procedure for total chlorine requires the reading to be made 2 minutes after the reagents are added pH Test Procedure 1 Remove one 1 eXact Strip Micro PH part number 486639 from the bottle before beginning the test Se...

Страница 7: ...n the display will show all annunciators then the current MENU selection followed by the last reading TURN METER ON 3 Press and re press the MENU button until the display shows the parameter AL4 SELECT TEST AL4 4 Rinse the CELL at least 3 times with the water sample you will be testing rinsing minimizes the potential for cross contamination from a previous test Finally fill cell to capacity 4ml wi...

Страница 8: ...TER WITH SAMPLE Copper Test Procedure CA5 Calcium reacts with Oxalic acid as it is released from the strip to form a white precipitate directly proportional to the Calcium concentration Values are reported as Calcium Carbonate CU6 Copper reacts with Biquinoline or Bicinchoninic Acid as it is released from the strip to form a purple color directly proportional to the copper concentration CA CA5 CU ...

Страница 9: ... charts round off your T value to two digits Tests that can be used with the Micro 7 in TR7 MENU are listed on page 31 Typically the most common tests are supplied with the conversion chart in this booklet For conversion charts and information about the tests not given in this booklet visit our website You can also use the TR7 MENU to develop your own custom conversion chart for unusual samples to...

Страница 10: ...TER WITH DISTILLED OR DEIONIZED WATER Turbidity Test Procedure TU TR7 Alkali pH Table Alkali pH results require the table below Follow eXact Micro 7 Alkali pH Test Procedure above using eXact Strip Micro Alkali pH Part No 486609 eXact Strip Micro Alkali pH Part No 486609 for 4mL Samples Rev 020509 AlkpH tr 90 80 70 60 50 40 30 20 10 0 9 8 7 6 5 4 3 2 1 0 8 3 8 3 8 3 8 3 8 5 8 7 8 9 9 1 9 3 9 6 8 3...

Страница 11: ...14 18 23 29 39 60 5 5 8 11 15 19 24 30 40 60 5 5 8 12 15 19 24 31 42 60 5 6 9 12 15 20 25 31 45 60 Cyanuric Acid Test Procedure CY TR7 Iron as Fe 2 Table Iron results require the table below Follow eXact Micro 7 Iron as Fe 2 Test Procedure above using eXact Strip Micro FE Part No 486631 6 Find the TR result in the table below to determine the Iron concentration in ppm parts per million Example a T...

Страница 12: ...0 03 0 03 0 39 1 06 1 84 2 8 4 1 5 8 5 8 5 8 12 NH3 TR7 Chlorine Dioxide Table Chlorine Dioxide results require the table below Follow eXact Micro 7 Chlorine Dioxide Test Procedure above using eXact Strip Micro CL DPD 1 Part No 486637 eXact Strip Micro CL DPD 1 Part No 486637 for 4mL Samples Rev 121908 ClO2 tr 90 80 70 60 50 40 30 20 10 0 9 8 7 6 5 4 3 2 1 0 0 03 0 23 0 51 0 82 1 19 1 62 2 2 2 9 4...

Страница 13: ...hloride as NaCl Test Procedure For Tap Water CHLORIDE TR7 6 Dip the eXact Strip Micro CHLORIDE as NaCl Part No 481657 into the CELL and immediately press READ This starts the 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion Remove and discard the strip when the time displays 1 The cursor will move across the display informing you that it is about to measu...

Страница 14: ...ples with suspended solids and particulates may cause incorrect results since the reducing agent in the powder pillow often dissolves the suspended solids and particulates If the meter is zeroed with the original water sample the zero is no longer valid for this kind of sample So for this kind of sample zero the meter using a water sample free of suspended solids and particulates use distilled dei...

Страница 15: ... 7 Find the TR result in the table below to determine the Manganese concentration in ppm parts per million Example a TR result of 65 3 use only the 65 for the chart equals a Manganese value of 0 04 ppm Record result After testing is completed rinse cell immediately USE TABLE 4 Press the ZERO ON button The cursor will move across the display followed by 100 T Sample is ready for testing ZERO METER ...

Страница 16: ...8 39 62 80 100 121 140 140 140 140 8 41 64 82 102 124 140 140 140 140 8 43 66 84 104 126 140 140 140 140 13 46 68 86 106 128 140 140 140 140 18 49 70 88 108 131 140 140 140 140 NOTE Divide the above Nitrate result by 4 4 to determine Nitrate value as Nitrogen NO3 as N 16 NO3 TR7 Nitrite NO2 Table Nitrite results require the table below Follow eXact Micro 7 Nitrite NO2 Test Procedure above using eX...