Calcium Hardness Test Procedure
1
Remove one (1)
eXact
®
Strip Micro CA
, part number 486629, from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the
eXact
®
Strip Micro CA, Part No. 486629
into the
CELL
and immediately press
READ
.
This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle back
and forth motion.
Remove and discard the strip when the time displays 1
. The cursor will
move across the display, informing you that it is about to measure the sample. Record result
displayed (this result is automatically stored in CA5). After testing is completed, rinse cell
immediately.
DIP STRIP AND PRESS “READ”
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 PPM
.
The sample is ready for testing.
ZERO METER
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the
MENU
button until the display shows the parameter
CA5
.
SELECT TEST: CA5
4
Rinse the
CELL
2 or 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml)
with the water sample.
FILL METER WITH SAMPLE
1
Remove one (1)
eXact
®
Strip Micro CU
, part number 486632, from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the
eXact
®
Strip Micro CU, Part No. 486632
into the
CELL
and immediately press
READ
.
This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle back
and forth motion.
Remove and discard the strip when the time displays 1
. The display will
flash
“SIt”
and begin immediately counting up from
1 to 20
(this extra time allows more thorough
color development). When the timer reaches
20
. At 20 sec, the cursor will move across the
display,
informing you that it is about to measure the sample. Record result displayed (this result
is automatically stored in CU6). After testing is completed, rinse cell immediately.
DIP STRIP AND PRESS “READ”
5
Press the
ZERO/ON
button. The cursor will move across the display, followed by
0.00 PPM
.
The sample is ready for testing.
ZERO METER
2
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the
MENU
button until the display shows the parameter
CU6
.
SELECT TEST: CU6
4
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing
minimizes the potential for cross-contamination from a previous test. Finally, fill cell to
capacity (4ml) with the water sample.
FILL METER WITH SAMPLE
Copper Test Procedure
CA5: Calcium reacts with Oxalic acid as it is released from the strip to form a white precipitate, directly
proportional to the Calcium concentration. Values are reported as Calcium Carbonate.
CU6: Copper reacts with Biquinoline or Bicinchoninic Acid as it is released from the strip to form a purple color,
directly proportional to the copper concentration.
CA
CA5
CU
CU6
Chromate (Chromium) Table
Nitrite results require the table below. Follow
eXact
®
Micro 7+ Chromate (Chromium) Test Procedure
(above)
using
eXact
®
Strip Micro CR, Part No. 486614.
eXact
®
Strip Micro CR Part No. 486614 - for 4mL Samples
Rev. 122908 CrO4
tr
90
80
70
60
50
40
30
20
10
0
9
8
7
6
5
4
3
2
1
0
<0.02
0.14
0.29
0.47
0.66
0.89
1.17
1.54
2.09
3.69
0.02
0.16
0.31
0.48
0.68
0.92
1.20
1.58
2.17
3.95
5
Dip the
eXact
®
Strip Micro CR, Part No. 486614
into the
CELL
and immediately press
READ
.
This starts the
20 SECOND
countdown timer. During this time move the strip in a gentle back
and forth motion.
Remove and discard the strip when the time displays 1
. The cursor will
move across the display, informing you that it is about to measure the sample (ignore this
result). Time the reaction in the cell for
three (3) minutes
(timer not included). During this time,
meter will shut off. When
3
minutes have elapsed, turn on the meter and wait for the last reading
to be displayed and then press
READ
, which will start a final
20 SECOND
countdown. The
cursor will move across the display, informing you that it is about to measure the sample.
Record result displayed (this result is automatically stored in TR7).
DIP STRIP -
(read carefully and follow procedure closely)
6
Find the “TR” result in the table below to determine the Chromate concentration in ppm (parts
per million). (Example: a “TR” result of 66.3 (use only the 66 for the chart) equals a Chromate
value of 0.52 ppm). Record result. After testing is completed, rinse cell immediately.
USE TABLE
4
Press the
ZERO/ON
button. The cursor will move across the display, followed by
100 %T
.
Sample is ready for testing.
ZERO METER
1
Press the
ZERO/ON
button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
2
Press and re-press the
MENU
button until the display shows the parameter TR7.
SELECT TEST: TR7
3
Rinse the
CELL
at least 3 times with the water sample you will be testing - rinsing minimizes
the potential for cross-contamination from a previous test. Finally, fill cell to capacity (4ml) with
the water sample.
FILL METER WITH SAMPLE
Chromate (Chromium) Test Procedure
0.03
0.17
0.33
0.50
0.70
0.94
1.23
1.63
2.26
4.16
0.04
0.19
0.35
0.52
0.73
0.97
1.27
1.67
2.35
4.36
0.06
0.20
0.36
0.54
0.75
0.99
1.30
1.72
2.46
4.55
0.07
0.22
0.38
0.56
0.77
1.02
1.34
1.77
2.57
4.72
0.09
0.23
0.40
0.58
0.80
1.05
1.37
1.83
2.72
4.89
0.10
0.25
0.41
0.60
0.82
1.08
1.41
1.89
2.89
>4.89
0.12
0.26
0.43
0.62
0.84
1.11
1.45
1.95
3.08
>4.89
0.13
0.28
0.45
0.64
0.87
1.14
1.49
2.02
3.36
>4.90
25
Cr
TR7